Getting Started
Rotary Switch
2
2-19
Setup
Allows you to do the following:
•
Select a test standard and cable type.
•
Select an average cable temperature when temperature is required by the
selected test standard.
•
Set the test tool to test cable installed in conduit when a conduit setting is
required by the selected test standard.
•
Enable remote testing or automatic remote detection when you use a second
main unit or a smart remote as a remote unit.
•
Set the cable identification number to increment automatically each time you
save Autotest results.
•
Set the backlight timer to turn off the backlight after a specified period of
inactivity.
•
Set the power-down timer to switch the test tool to a low-power mode after a
specified period of inactivity.
•
Set the fault threshold for the impulse noise test.
•
Select interface parameters for the serial port.
•
Enable or disable the test tool’s beeper.
•
Set the date and time.
•
Select a format for the date and time.
•
Select a unit for length measurements.
•
Select a format for displaying decimal fractions.
•
Select a language for the display and printed reports.
•
Select a frequency for the power line noise filter.
•
Enable or disable the shield continuity test.
•
Modify test standards for custom cable configurations.
•
Select 100 MHz or 155 MHz as the maximum frequency for NEXT, ACR, and
attenuation tests.