Tracking Generator Specifications

(Option TG3 required)

Supplemental information

Output frequency

Range

100 kHz to 3 GHz

Settable to 9 kHz

Resolution

1 Hz

Output power level

Range

-30 to 0 dBm

Resolution

0.1 dB

Absolute accuracy

± 0.75 dB

20 to 30 °C, at 50 MHz with coupled source attenuator,
referenced to -20 dBm

Output flatness

± 3 dB

100 kHz to 10 MHz

± 2 dB

10 MHz to 3 GHz

VSWR

< 1.5:1

300 kHz to 3 GHz, input attenuator ≥ 12 dB

Connector and impedance

N-type female, 50 Ω

Maximum safe reverse level

Average total power

30 dBm (1 W)

AC coupled

0 VDC MAX

Supplemental information

Spurious response (continued)

Input related spurious

< -60 dBc

-30 dBm signal at input mixer, 20 to 30 °C

Residual response (inherent)

< -83 dBc

Input terminated and 0 dB RF attenuation, preamp off

Phase noise

Specification

Typical

Offset from CW signal

10 kHz

< -88 dBc/Hz

< -90 dBc/Hz

Fc = 1 GHz, RBW = 1 kHz, VBW = 10 Hz,
and sample detector, log average,
average times > 40

100 kHz

< -100 dBc/Hz

< -102 dBc/Hz

1 MHz

< -110 dBc/Hz

< -112 dBc/Hz

Residual FM

≤ 100 Hz peak-to-peak
in 100 ms

1 kHz RBW, 1 kHz VBW

Dynamic Range Specifications