beautypg.com

Amplitude specifications, Continued) – Atec Agilent-N9320B User Manual

Page 5

background image

5

Supplemental information

Level measurement uncertainty

20 to -30 °C; frequency > 1 MHz; signal input 0 to -50 dBm; reference level 0 to -50 dBm; input attenuation 20 dB; RBW
1 kHz, VBW 1 kHz; after calibration; preamp off

Overall amplitude accuracy

± 1.5 dB

± 0.5 dB, typical

Level display range

Log scale units

dBm, dBmV, dBμV, dBμA

Linear scale and units

W, mW, μW, A, mA, μA, V, mV, μV

Marker level readout

0.01 dB

Resolution

0.01% of reference level

Log scale

Number of traces

4

Linear scale

Detectors

Positive-peak, negative-peak, sample,
normal, RMS

Trace functions

Clear/write, maximum hold, average,
minimum hold, view

Preamplifier

Frequency range

1 MHz to 3.0 GHz

Gain

18 dB nominal

Amplitude Specifications

(continued)