beautypg.com

Frequency specifications – Atec Agilent-E4408B User Manual

Page 4

background image

Frequency Specifications

a.

N is the harmonic mixing mode. For negative mixing modes (as indicated by "-"), the desired first LO harmonic is higher than the tuned frequency by the first IF

(3.9214 for the 9 kHz to 3 GHz band, and 321.4 MHz for all other bands.)

b.

+5% of span + . Sweep points fixed at 401 for basic analyzer.

c.

Not available in RBW < 1 kHz (Option 1DR).

span

sweep pts. – 1

4

E4411B

Frequency range

E4403B

E4408B

BAS/BTG configuration

9 kHz - 1.5 GHz

9 kHz - 3 GHz

9 kHz - 26.5 GHz

Custom configuration

N/A

N/A

(75 Ω input Option 1DP)

1 MHz - 1.5 GHz

Frequency range

100 Hz - 3 GHz

2.85 - 6.7 GHz

6.2 - 13.2 GHz

12.8 - 19.2 GHz

18.7 - 26.5 GHz

Band

0

1

2

3

4

Harmonic (N

a

)

mixing mode

1-

1-

2-

4-

4-

Logarithmic scale

N/A

Range = 0 Hz (zero span), 100 Hz to maximum frequency range of the analyzer

Accuracy

Linear scale

1% of span

Marker frequency counter

c

Accuracy = ± (marker frequency x frequency reference error + counter resolution)

Counter resolution = selectable from 1 Hz to 100 kHz

Frequency span

Span coefficient (SP)

b

0.75% x span

External reference

10 MHz

Temperature stability

–6

Settability

–7

Frequency readout accuracy (start, stop, center, marker)

= ± (frequency indication x frequency reference error + SP

b

+15% of RBW + 10 Hz + 1 Hz x N

a

)

Aging rate

–6

±5 x 10 /year

±5 x 10 /year

±2 x 10 /year

Frequency reference

Frequency reference error = ± [(aging rate x time since last adjustment ) + settability + temperature stability]

Basic analyzer