Specifications, Cw mode, Analog sweep mode – Atec Anritsu-68369ANV User Manual

Page 8: Phase-locked step sweep mode, Alternate sweep mode, Manual sweep mode, List sweep mode, Programmable frequency agility, Markers, Sweep triggering

background image

6

CW Mode

Output:

Twenty independent, presettable CW frequencies

(F0 – F9 and M0 – M9).

Accuracy:

Same as internal or external 10 MHz time base.

Internal Time Base Stability:

with aging:

< 2 x 10

-8

/day (

<5 x 10

–10

/day with Option 16)

with temperature:

< 2 x 10

-8

/

°C over 0°C to 55°C

(

<2 x 10

-10

/

°C with Option 16)

Resolution:

1 kHz (0.1 Hz with Option 11)

External 10 MHz Reference Input:

Accepts external 10 MHz

±100 Hz, –10 to +20 dBm time base signal. Automatically
disconnects the internal high-stability time-base option, if installed
BNC, rear panel, 50

Ω impedance.

10 MHz Reference Output:

0 dBm into 50

Ω, AC coupled, from

rear panel BNC connector.

Switching Time (typical maximum):

Units with maximum frequency

20 GHz: <40 ms to be within

1 kHz of final frequency

Units with maximum frequency of 8.4 GHz: <15 ms to within

1 kHz of final frequency

Analog Sweep Mode

(681XXC, 683XXC, 691XXB and 693XXB only)

Sweep Width:

Independently selected from 1 MHz to full

frequency range.

Accuracy:

The lesser of:

±30 MHz or (±2 MHz + 0.25% of sweep width) for

sweep speeds of

≤50 MHz/ms

Sweep Time Range:

30 ms to 99 seconds

Phase-Locked Step Sweep Mode

Sweep Width:

Independently selected, 1 kHz (0.1 Hz with

Option 11) to full range.

Accuracy:

Same as internal or external 10 MHz time base.

Resolution (Minimum Step Size):

1 kHz (0.1 Hz with Option 11)

Number of Steps:

Variable from 1 to 10,000

Step Size:

1 kHz (0.1 Hz with Option 11) to the full frequency range

of the instrument. (If the step size does not divide into the
selected frequency range, the last step is truncated.)

Step Time:

Step Sweep: Variable from 1 ms/step to 99 seconds/step.

Dwell time begins after phase lock.

Fixed Rate Step Sweep: Variable from 20 ms/step to

99 seconds/step. Dwell time includes phase lock time.

Switching Time (typical maximum):

Units having a high-end frequency of

20 GHz:

<15 ms + 1 ms/GHz step size or <40 ms, whichever is less.

Units having a high-end frequency of 8.4 GHz: < 7 ms

Alternate Sweep Mode

Sweeps alternately between any two sweep ranges. Each
sweep range may be associated with a different power level.

Manual Sweep Mode

Provides stepped, phase-locked adjustment of frequency
between sweep limits. User-selectable number of steps or
step size.

List Sweep Mode

Under GPIB control or via the front panel, up to 4 tables with
2000 non-sequential frequency/power sets can be stored and
then addressed as a phase-locked step sweep. One table of
2000 points is stored in non-volatile memory, all other tables
are stored in volatile memory.

Switching Time (typical maximum):

Units having a high-end frequency of

20 GHz: <25 ms to be

within 1 kHz of final frequency.

Units having a high-end frequency of 8.4 GHz: < 5 ms to be

within 1 kHz of final frequency.

Programmable Frequency Agility

Under GPIB control, up to 3202 non-sequential frequency/
power sets can be stored and then addressed as a phase-
locked step sweep. Data is stored in volatile memory.

Markers

Up to 20 independent, settable markers (F0 – F9 and
M0 – M9).

Video Markers:

+5V or –5V marker output, selectable from system

menus. AUX I/O connector, rear panel.

Intensity Markers (Available in analog sweeps of

<1 second sweep time):

Produces an intensified dot on trace,

obtained by momentary dwell in RF sweep.

Marker Accuracy:

Same as sweep frequency accuracy.

Marker Resolution :

Analog Sweep: 1 MHz or Sweep Width/4096,

whichever is greater.

Step Sweep: 1 kHz (0.1 Hz with Option 11)

Sweep Triggering

Sweep triggering is provided for Analog Frequency Sweep if
applicable, Step Frequency Sweep, List Frequency Sweep,
and CW Power Sweep.

Auto:

Triggers sweep automatically.

External:

Accepts a TTL low to high transition at AUX I/O connector

on rear panel to trigger a sweep.

Single:

Triggers, aborts, and resets a single sweep. Reset sweep

may be selected to be at the top or bottom of the sweep. The
68100C/300C and 69100B/300B pen lift will activate at sweep
times

≥1 second.

SPECIFICATIONS