Amplitude Modulation

All amplitude modulation specifications apply at 50% depth,
1 kHz rate, with RF level set 6 dB below maximum specified
leveled output power, unless otherwise noted.

AM Depth (typical):

0-90% linear; 20 dB log

AM Bandwidth (3 dB):

DC to 50 kHz minimum
DC to 100 kHz typical

Flatness (DC to 10 kHz rates):

± 0.3 dB

Accuracy:

± 5%

Distortion:

< 5% typical

Incidental Phase Modulation
(30% depth, 10 kHz rate):

< 0.2 radians typical

External AM Input:

Log AM or Linear AM input, front or rear-panel

BNC, 50

Ω or 600 Ω input impedance. All options selectable from

modulation menu.
Sensitivity:

Log AM: Continuously variable from 0 dB per volt to

25 dB per volt.

Linear AM: Continuously variable from 0% per volt

to 100% per volt.

Maximum Input: ±1V

Internal AM Generator:

Waveforms: Sinusoid, squarewave, triangle, positive ramp,

negative ramp, Gaussian noise, uniform noise, user-defined➃

Rate: 0.1 Hz to 1 MHz sinusoidal

0.1 Hz to 100 kHz squarewave, triangle, ramps

Resolution: 0.1 Hz
Accuracy: Same as instrument timebase
Output: BNC connector, rear panel

Frequency Modulation

Maximum FM Deviation:

Locked Mode (1 kHz to 8 MHz rates): The lesser of ±10 MHz or

rate x 300
Locked Low Noise Mode (50 kHz to 8 MHz rates): The lesser of

±10 MHz or rate x 3

Unlocked Narrow Mode (DC to 8 MHz rates): ±10 MHz
Unlocked Wide Mode (DC to 100 Hz rates): ±100 MHz

FM Bandwidth (3 dB):

Locked Mode: 1 kHz to 10 MHz
Locked Low Noise Mode: 30 kHz to 10 MHz
Unlocked Narrow Mode: DC to 10 MHz
Unlocked Wide Mode: DC to 100 Hz

Flatness (locked mode 10 kHz to 1 MHz rates):

±1 dB

Accuracy (100 kHz rate):

10% (5% typical)

Incidental AM (± 1 MHz deviation, 1 MHz rate): < 2%

Harmonic Distortion (±1 MHz deviation, 10 kHz rate):

<1%

External FM Input:

Front or rear panel BNC, 50

Ω or 600 Ω input

impedance. All options selectable from modulation menu.
FM Sensitivity: Continuously variable from ±10 kHz per volt to

±20 MHz per volt (Locked, Locked Low Noise and Unlocked
Narrow FM modes), or ±100 kHz per volt to ±100 MHz per volt
(Unlocked Wide FM mode), selectable from modulation menu.

Maximum Input: ± 1V

Internal FM Generator:

Waveforms: Sinusoid, squarewave, triangle, positive ramp,

negative ramp, Gaussain noise, uniform noise, user-defined

Rate: 0.1 Hz to 1 MHz sinusoidal

0.1 Hz to 100 kHz squarewave, triangle, ramps

Resolution: 0.1 Hz
Accuracy: Same as instrument timebase
Output: BNC connector, rear panel

Phase Modulation (FM, Option 6)

ΦM Deviation:

Narrow Mode (DC to 8 MHz Rates): The lesser of ±3 radians

or ±5 MHz/rate

Wide Mode (DC to 1 MHz Rates): The lesser of ±400 radians

or ±10 MHz/rate.

ΦM Bandwidth (3 dB, relative to 100 kHz rate):

Narrow Mode: DC to 10 MHz
Wide Mode: DC to 1 MHz

ΦM Flatness (relative to 100 kHz rate):

Narrow Mode (DC to 1 MHz rates): ±1 dB
Wide Mode (DC to 500 kHz rates): ±1 dB

ΦM Accuracy (at 100 kHz sine wave): 10%

External

ΦM Input

: Front or rear panel BNC (shares the

FM input), 50

Ω or 600 Ω input impedance. All options selectable

from modulation menu. Shares connectors with FM.
ΦM Sensitivity: Continuously variable from ±0.0025 radians per

volt to ±5.0 radians per volt (Narrow

ΦM mode) or ±0.25

radians per volt to ± 500.0 radians per volt (Wide

ΦM mode),

selectable from modulation menu.

683XXC/693XXB MODULATION

➃ User-defined waveforms are available with Option 10.

➄ For external input, accuracy applies at ±1 V input