beautypg.com

Other output power specifications, Cw power sweep, Sweep frequency/step power – Atec Anritsu-68369ANV User Manual

Page 13: Internal power measurement

background image

11

➂ When Option 8 Power Meter is installed, Option 7 (Delete AM/FM

Generators) is not available.

Other Output Power Specifications

Output Units:

Output units selectable as either dBm or mV.

Selection of mV assumes 50

Ω load. All data entries and displays are

in the selected units.

Output Power Resolution:

0.01 dB or 0.001 mV

Source Impedance:

50

Ω nominal

Source SWR (Internal Leveling):

<2.0 typical

Power Level Stability with Temperature:

0.04 dB/

°C typical

Level Offset:

Offsets the displayed power level to establish a new

reference level.

Internal Leveling:

Power is leveled at the output connector in

all modes.

External Leveling:

External Detector: Levels output power at a remote detector
location. Accepts a positive or negative 0.5 mV to 500 mV input
from the remote detector. EXT ALC ADJUST adjusts the input
signal range to an optimum value. BNC connector, front and
rear panel.
External Power Meter: Levels output power at a remote power
meter location. Accepts a ± 1V full scale signal from the remote
power meter. EXT ALC ADJUST adjusts the input signal range to
an optimum value. BNC connector, front and rear panel.
External Leveling Bandwidth:

30 kHz typical in Detector mode.
0.7 Hz typical in Power Meter mode.

User Level Flatness Correction:

Number of points: 2 to 801 points per table
Number of tables: 5 available
Entry modes: GPIB power meter or computed data

CW Power Sweep

Range:

Sweeps between any two power levels at a single

CW frequency.

Resolution:

0.01 dB/step (Log) or 0.001 mV/step (Linear)

Accuracy:

Same as CW power accuracy.

Step Size:

User-controlled, 0.01 dB (Log) or 0.001 mV (Linear) to

the full power range of the instrument.

Step Dwell Time:

Variable from 1 ms to 99 seconds. If the sweep

crosses a step attenuator setting, there will be a sweep dwell of
approximately 20 ms to allow setting of the step attenuator.

Sweep Frequency/Step Power

A power level step occurs after each frequency sweep. The
power level remains constant for the length of time required
to complete each sweep.

Internal Power Measurement

(Option 8 on 68300C and 69300B only)

Sensors:

Compatible with Anritsu 560-7, 5400-71, or

6400-71 Series Detectors. Rear panel input.

Range:

+16 dBm to

−35 dBm.

Accuracy:

±1 dB (+10 dBm to –10 dBm)
± 2 dB (-10 dBm to –35 dBm)

Resolution:

0.1 dB minimum