beautypg.com

Pulse modulation, Scan modulation (option 20) – Atec Anritsu-68369ANV User Manual

Page 16

background image

Internal

ΦM Generator

(Shares the Internal FM Generator)

Waveforms: Sine, square, triangle, positive ramp, negative ramp,

Gaussian noise, uniform noise, user defined

Rate: 0.1 Hz to 1 MHz for sine wave

0.1 Hz to 100 kHz for other waveforms

Resolution: 0.1 Hz
Accuracy: Same as instrument timebase.
Output: BNC connector, rear panel

Pulse Modulation

Pulse modulation specifications apply at maximum rated
power, unless otherwise noted.

On/Off Ratio:

>80 dB

Rise/Fall Time (10 to 90%):

10 MHz to 1.0 GHz: <15 ns (<10 ns typical)
1.0 GHz to 65 GHz: <10 ns (<5 ns typical)

Minimum Leveled Pulse Width:

<100 ns,

≥2 GHz

<1

µs, <2 GHz

Minimum Unleveled Pulse Width:

<10 ns

Pulse Overshoot:

<10%➆

Level Accuracy Relative to CW
(100 Hz to 1 MHz PRF):

±0.5 dB,

≥1 µs pulse width

±1.0 dB,

<1 µs pulse width

Video Feedthrough:

< ±10 mV, ≥2 GHz

Pulse Width Compression:

< 8 ns typical

Pulse Delay (typical):

PRF Range:

DC to 10 MHz unleveled
100 Hz to 5 MHz leveled

External Input:

Front or rear-panel BNC, selectable from

modulation menu.
Drive Level: TTL compatible input
Input Logic: Positive-true or negative-true, selectable from

modulation menu.

Internal Pulse Generator:

Modes:

Free-run, triggered, gated, delayed, singlet, doublet,

triplet, quadruplet

Accuracy: 10 ns (5 ns typical)
Inputs/Outputs:

Video pulse and sync out, rear-panel

BNC connectors

Scan Modulation (Option 20)

Frequency Range:

1 to 20 GHz

Attenuation Range:

0 to 60 dB

Flatness:

±2 dB, 0 to 40 dB
±3.5 dB, 40 to 60 dB

Step Response:

<1

µs

Sensitivity:

−10 dB/V

Insertion Loss
(when engaged):

<6 dB, 1 to 18 GHz
<8 dB, 18 to 20 GHz

Input:

Rear-panel BNC (f) connector

14

683XXC/693XXB MODULATION

➅ User-defined waveforms are available with Option 10.

➆ For 50, 60 and 65 GHz units, overshoot from 40 to 65 GHz is 20% typical

at rated power.

➇ Period must be longer than the sum of delay and width by 5 clock

cycles minimum.

➈ Option 20, SCAN Modulator is available on models 68337C, 68347C, 69337B

and 69347B only. When Option 20 SCAN Modulator is installed, Option 7 and
Option 8 are not available.

Mode

Pulse Delay (ns)

External

50

Triggered

100

Triggered with Delay

200

Parameter

Selectable Clock Rate

40 MHz

10 MHz

Pulse Width

25 ns to 419 ms

100 ns to 1.6s

Pulse Period

250 ns to 419 ms

600 ns to 1.6s

Variable Delay

Singlet

Doublet

Triplet

Quadruplet

0 to 419 ms

100 ns to 419 ms
100 ns to 419 ms
100 ns to 419 ms

0 to 1.6s

300 ns to 1.6s
300 ns to 1.6s
300 ns to 1.6s

Resolution

25 ns

100 ns