beautypg.com

Atec Agilent-81180A User Manual

Page 17

background image

17

Trigger input

Characteristics

Description

Connector type

SMA

Drive

Channel 1, channel 2, or both

Input impedance

10 kΩ, typical

Polarity

Positive, negative, or both, selectable

Damage level

± 20 Vdc

Frequency range

0 to 15 MHz

Trigger level control
Range

-5 V to 5 V

Resolution

12 bit (2.5 mV)

Accuracy

± (5% of setting + 2.5 mV)

Sensitivity

200 mVp-p

Pulse width, minimum

10 ns

System delay

200 sample clock periods + 50 ns, typical

Trigger delay

Separate for each channel

Range

0 to 8,000,000 sample clock periods

Resolution

8 points

Accuracy

Same as sample clock accuracy

Smart trigger

Detects a unique pulse width range

Conditioned trigger

< pulse width, > pulse width, <> pulse width

Pulse width range

50 ns to 2 s

Resolution

2 ns

Accuracy

± (5% of setting +20 ns)

Trigger holdoff

Ignores triggers for a holdoff duration

Holdoff range

100 ns to 2 s

Resolution

2 ns

Accuracy

± (5% of setting +20 ns)

Trigger jitter

8 sampling periods

Internal trigger generator

Characteristics

Description

Source

Common or separate for each channel

Mode

Timer (waveform start to waveform start); delayed (waveform stop to waveform start)

Timer
Range

100 ns to 2 s

Resolution

3 digits

Accuracy

100 ppm

Delayed
Range

152 to 8,000,000 sample clock periods

Resolution

Integer numbers, divisible by 8

4. Trigger input to analog output