Atec Agilent-81180A User Manual
Page 17

17
Trigger input
Characteristics
Description
Connector type
SMA
Drive
Channel 1, channel 2, or both
Input impedance
10 kΩ, typical
Polarity
Positive, negative, or both, selectable
Damage level
± 20 Vdc
Frequency range
0 to 15 MHz
Trigger level control
Range
-5 V to 5 V
Resolution
12 bit (2.5 mV)
Accuracy
± (5% of setting + 2.5 mV)
Sensitivity
200 mVp-p
Pulse width, minimum
10 ns
System delay
200 sample clock periods + 50 ns, typical
Trigger delay
Separate for each channel
Range
0 to 8,000,000 sample clock periods
Resolution
8 points
Accuracy
Same as sample clock accuracy
Smart trigger
Detects a unique pulse width range
Conditioned trigger
< pulse width, > pulse width, <> pulse width
Pulse width range
50 ns to 2 s
Resolution
2 ns
Accuracy
± (5% of setting +20 ns)
Trigger holdoff
Ignores triggers for a holdoff duration
Holdoff range
100 ns to 2 s
Resolution
2 ns
Accuracy
± (5% of setting +20 ns)
Trigger jitter
8 sampling periods
Internal trigger generator
Characteristics
Description
Source
Common or separate for each channel
Mode
Timer (waveform start to waveform start); delayed (waveform stop to waveform start)
Timer
Range
100 ns to 2 s
Resolution
3 digits
Accuracy
100 ppm
Delayed
Range
152 to 8,000,000 sample clock periods
Resolution
Integer numbers, divisible by 8
4. Trigger input to analog output