beautypg.com

Atec Agilent-81180A User Manual

Page 11

background image

11

Waveform type

Characteristics

Description

Standard

A waveform is selected from a built-in library. The standard waveform parameters are

programmable.

Arbitrary

Arbitrary waveform coordinates are downloaded and stored in memory segments. The

arbitrary waveform parameters are programmable.

Sequenced

Arbitrary waveforms are downloaded and stored in memory segments. The segments

are arranged in a sequence table that step, loop, jump and nest on segments in a user-

defined configuration. Conditional jump and nest pending an event signal.

Advanced sequences

Same functionality as described for sequenced waveforms except sequences are arranged

in the sequence table.

Modulated

A modulated waveform is calculated from a built-in library of modulation schemes.

Pulse

A pulse waveform is calculated and downloaded to the arbitrary waveform memory.

Run mode

Characteristics

Description

Continuous
Self armed

A selected output function shape is output continuously. No start commands are

required to generate waveforms.

Armed

No start commands are required to generate waveforms.

Triggered

A trigger signal activates a single-shot or counted burst of output waveforms and then

the instrument waits for the next trigger signal.

Normal mode

The first trigger signal activates the output; consecutive triggers are ignored for the

duration of the output waveform.

Override mode

The first trigger signal activates the output; consecutive triggers restart the output

waveform whether the current waveform has been completed or not.

Gated

A waveform is output when a gate signal is asserted. The waveform is repeated until

the gate signal is de-asserted. Last period is always completed.

Standard waveforms

Characteristics

Description

General

Waveforms are computed and generated every time a standard waveform is selected.

Standard waveform library

Built-in, auto computed waveforms: sine, triangle, square, ramp, pulse, sink,

exponential rise, exponential decay, Gaussian, noise and DC.

Standard waveform control

The standard waveform parameters can be adjusted to specific requirements. The

waveform is recomputed with each parameter change.