beautypg.com

Atec Anritsu-MG3670 Series User Manual

Page 5

background image

RADIO COMMUNICATIONS, EMI MEASURING TEST INSTRUMENTS

294

www.anritsu.com

For product ordering information, see pages 4-7.

Internal modulation

Depends on installed modulation unit (MG0301C/0302A/0305A/0307A/0310A/0311A/0312A/0314A)

For MG3670B/C, MG3671A/B

Any modulation using I/Q input signal
Input frequency: DC to 1.2 MHz*

3

Input level:

I

2

± Q

2

≤0.5 Vrms, BNC connector

*

I/Q

≤1.5 Vp-p (50 Ω), I/Q ≤10% to 100% of 1.5 Vp-p (CMOS)

Digital

External

Vector error:

≤1.8%rms (I/Q input level: 1 Vrms/50 Ω, at ≤+5 dBm output)

modulation

modulation

For MG3672A

50

Ω input

Input frequency: DC to 30 MHz (BW: 3 dB, 18˚ to 30˚C), Input level:

I

2

± Q

2

≤1.0 Vrms, I/Q ≤1.5 Vp-p

CMOS input

Input frequency: DC to 1.2 MHz, Input level:

I

2

± Q

2

≤1.0 Vrms, I/Q ≤1.5 Vp-p

I/Q output

Outputs I/Q signal at internal modulation (MG0301C/0302A/0305A/0307A/0310A/0311A/0312A/0314A installed)

Pulse

Input

TTL level, BNC connector, polarity selectable

modulation

On/off ratio

≥40 dB (at ≥0 dBm output)

Transition time

≤2 µs, minimum pulse width: 10 µs

Memory

Frequency memory

1000 carrier frequencies (save and recall)

function

Parameter memory

100 panel settings (save and recall)

Relative display

Carrier frequency, output level

I/Q signal adjustment

Variable offset, balance, phase (only output) of I/Q input/output signal (DC to 1.2 MHz)

Other

Backup

Last settings stored at power-off

functions

Reverse power
protection

Maximum reverse input power: 50 W (<1000 MHz), 25 W (

≥1000 MHz), ±50 Vdc

GPIB

All functions except power switch and panel lock switch controlled
Interface function: SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT0, C0, E2

Operating temperature

0˚ to 50˚C

Power

100 to 120/200 to 240 Vac (switchable), 47.5 to 63 Hz,

≤550 VA

Dimensions and mass

(426±5) W x (221.5±4) H x (451±5) D mm,

≤27 kg

EMC*

4

EN55011 (1991, Group 1, Class A), EN50082-1 (1992), Harmonic current emissions EN61000-3-2 (1995)

Safety

EN61010-1: 1993 (Installation Category

ΙΙ, Pollution Degree ΙΙ)

*

1: Internal reference oscillator accuracy: 2 x 10

–8

/day (23˚ ±5˚C), calibrated after 24 h operation

*

2: Depended on the specifications of each units when MG0310A or MG0314A unit are installed.

*

3: Refer to the “Frequency response for I/Q external modulation (typical value)” on page 262 for the input frequency range. Typical values are given for refer-

ence only to assist in the use of this instrument, and are not guaranteed specifications.

*

4: Electromagnetic compatibility

MG0301C

π/4 DQPSK Modulation Unit (incorporated in the MG3670B/C, MG3671A/B and MG3672A)

Applicable
communication system

PDC, PDC_H, PHS, NADC, TFTS

Modulation system

π/4 DQPSK

Vector error

I/Q signal:

≤1.5%rms (at 50 Ω output), RF signal: ≤1.8%rms (at ≤+5 dBm output)

Internal modulation

Pseudorandom pattern: PN15, PN9

data

Free 4-bit repetition pattern (ex: 1010, 1111)

DATA CLOCK: Covering ±5% of bit rate

External modulation

DATA: Digital data synchronized with DATA CLOCK

data

SYMBOL CLOCK: Clock specified by DATA synchronized with DATA CLOCK
TTL level, BNC connector, polarity selectable

Selectable 50

Ω or CMOS (600 Ω), BNC connector

50

Ω setting [modulation data: 0000 (TFTS: 1111)]: 1 Vp-p ±2% (MG3670A/B/C, MG3671A/B), 2 Vp-p ±2% (MG3672A)

CMOS setting [modulation data: 0000 (TFTS: 1111)]

I/Q signal output

Variable in 10% steps over range of 10% to 100% of 1 Vp-p ±2%, variable offset voltage: 0 to 4 V in 1 mV steps (MG3670A/B/C,
MG3671A/B)
Variable in 10% steps over range of 10% to 100% of 2 Vp-p ±2%, variable offset voltage: 0 to 4 V in 1 mV steps (MG3672A)

Carrier frequency range:

300 kHz to 2250 MHz*

1

(incorporated in the MG3670B/C), 300 kHz to 2750 MHz (incorporated in the MG3671A/B and MG3672A)

PDC, PDC_H

Bit rate: 42 kbps
Baseband filter: Root Nyquist (

α = 0.5), Nyquist (α = 0.5)

Carrier frequency range:

1 to 2250 MHz*

1

(incorporated in the MG3670B/C), 1 to 2750 MHz (incorporated in the MG3671A/B and MG3672A)

PHS

Bit rate: 384 kbps
Baseband filter: Root Nyquist (

α = 0.5), Nyquist (α = 0.5)

Adjacent channel power ratio:

≤–74 dB (600/900 kHz offset, ±96 kHz band, ≥10 MHz)*

2

Carrier frequency range:

300 kHz to 2250 MHz*

1

(incorporated in the MG3670B/C), 300 kHz to 2750 MHz (incorporated in the MG3671A/B and MG3672A)

NADC

Bit rate: 48.6 kbps
Baseband filter: Root Nyquist (

α = 0.35), Nyquist (α = 0.35)

Carrier frequency range:

300 kHz to 2250 MHz*

1

(incorporated in the MG3670B/C), 300 kHz to 2750 MHz (incorporated in the MG3671A/B and MG3672A)

TFTS

Bit rate: 44.2 kbps
Baseband filter: Root Nyquist (

α = 0.4), Nyquist (α = 0.4)

*

1: The upper frequency is limited by the specifications of the main frame in which this unit is installed.

*

2: Applicable when this unit is installed in MG3670B/C, MG3671A/B and MG3672A. Not applicable when this unit is installed in MG3670A.