Atec Anritsu-MG3670 Series User Manual
Page 5
RADIO COMMUNICATIONS, EMI MEASURING TEST INSTRUMENTS
294
www.anritsu.com
For product ordering information, see pages 4-7.
Internal modulation
Depends on installed modulation unit (MG0301C/0302A/0305A/0307A/0310A/0311A/0312A/0314A)
For MG3670B/C, MG3671A/B
Any modulation using I/Q input signal
Input frequency: DC to 1.2 MHz*
3
Input level:
√
I
2
± Q
2
≤0.5 Vrms, BNC connector
*
I/Q
≤1.5 Vp-p (50 Ω), I/Q ≤10% to 100% of 1.5 Vp-p (CMOS)
Digital
External
Vector error:
≤1.8%rms (I/Q input level: 1 Vrms/50 Ω, at ≤+5 dBm output)
modulation
modulation
For MG3672A
50
Ω input
Input frequency: DC to 30 MHz (BW: 3 dB, 18˚ to 30˚C), Input level:
√
I
2
± Q
2
≤1.0 Vrms, I/Q ≤1.5 Vp-p
CMOS input
Input frequency: DC to 1.2 MHz, Input level:
√
I
2
± Q
2
≤1.0 Vrms, I/Q ≤1.5 Vp-p
I/Q output
Outputs I/Q signal at internal modulation (MG0301C/0302A/0305A/0307A/0310A/0311A/0312A/0314A installed)
Pulse
Input
TTL level, BNC connector, polarity selectable
modulation
On/off ratio
≥40 dB (at ≥0 dBm output)
Transition time
≤2 µs, minimum pulse width: 10 µs
Memory
Frequency memory
1000 carrier frequencies (save and recall)
function
Parameter memory
100 panel settings (save and recall)
Relative display
Carrier frequency, output level
I/Q signal adjustment
Variable offset, balance, phase (only output) of I/Q input/output signal (DC to 1.2 MHz)
Other
Backup
Last settings stored at power-off
functions
Reverse power
protection
Maximum reverse input power: 50 W (<1000 MHz), 25 W (
≥1000 MHz), ±50 Vdc
GPIB
All functions except power switch and panel lock switch controlled
Interface function: SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT0, C0, E2
Operating temperature
0˚ to 50˚C
Power
100 to 120/200 to 240 Vac (switchable), 47.5 to 63 Hz,
≤550 VA
Dimensions and mass
(426±5) W x (221.5±4) H x (451±5) D mm,
≤27 kg
EMC*
4
EN55011 (1991, Group 1, Class A), EN50082-1 (1992), Harmonic current emissions EN61000-3-2 (1995)
Safety
EN61010-1: 1993 (Installation Category
ΙΙ, Pollution Degree ΙΙ)
*
1: Internal reference oscillator accuracy: 2 x 10
–8
/day (23˚ ±5˚C), calibrated after 24 h operation
*
2: Depended on the specifications of each units when MG0310A or MG0314A unit are installed.
*
3: Refer to the “Frequency response for I/Q external modulation (typical value)” on page 262 for the input frequency range. Typical values are given for refer-
ence only to assist in the use of this instrument, and are not guaranteed specifications.
*
4: Electromagnetic compatibility
MG0301C
π/4 DQPSK Modulation Unit (incorporated in the MG3670B/C, MG3671A/B and MG3672A)
Applicable
communication system
PDC, PDC_H, PHS, NADC, TFTS
Modulation system
π/4 DQPSK
Vector error
I/Q signal:
≤1.5%rms (at 50 Ω output), RF signal: ≤1.8%rms (at ≤+5 dBm output)
Internal modulation
Pseudorandom pattern: PN15, PN9
data
Free 4-bit repetition pattern (ex: 1010, 1111)
DATA CLOCK: Covering ±5% of bit rate
External modulation
DATA: Digital data synchronized with DATA CLOCK
data
SYMBOL CLOCK: Clock specified by DATA synchronized with DATA CLOCK
TTL level, BNC connector, polarity selectable
Selectable 50
Ω or CMOS (600 Ω), BNC connector
50
Ω setting [modulation data: 0000 (TFTS: 1111)]: 1 Vp-p ±2% (MG3670A/B/C, MG3671A/B), 2 Vp-p ±2% (MG3672A)
CMOS setting [modulation data: 0000 (TFTS: 1111)]
I/Q signal output
Variable in 10% steps over range of 10% to 100% of 1 Vp-p ±2%, variable offset voltage: 0 to 4 V in 1 mV steps (MG3670A/B/C,
MG3671A/B)
Variable in 10% steps over range of 10% to 100% of 2 Vp-p ±2%, variable offset voltage: 0 to 4 V in 1 mV steps (MG3672A)
Carrier frequency range:
300 kHz to 2250 MHz*
1
(incorporated in the MG3670B/C), 300 kHz to 2750 MHz (incorporated in the MG3671A/B and MG3672A)
PDC, PDC_H
Bit rate: 42 kbps
Baseband filter: Root Nyquist (
α = 0.5), Nyquist (α = 0.5)
Carrier frequency range:
1 to 2250 MHz*
1
(incorporated in the MG3670B/C), 1 to 2750 MHz (incorporated in the MG3671A/B and MG3672A)
PHS
Bit rate: 384 kbps
Baseband filter: Root Nyquist (
α = 0.5), Nyquist (α = 0.5)
Adjacent channel power ratio:
≤–74 dB (600/900 kHz offset, ±96 kHz band, ≥10 MHz)*
2
Carrier frequency range:
300 kHz to 2250 MHz*
1
(incorporated in the MG3670B/C), 300 kHz to 2750 MHz (incorporated in the MG3671A/B and MG3672A)
NADC
Bit rate: 48.6 kbps
Baseband filter: Root Nyquist (
α = 0.35), Nyquist (α = 0.35)
Carrier frequency range:
300 kHz to 2250 MHz*
1
(incorporated in the MG3670B/C), 300 kHz to 2750 MHz (incorporated in the MG3671A/B and MG3672A)
TFTS
Bit rate: 44.2 kbps
Baseband filter: Root Nyquist (
α = 0.4), Nyquist (α = 0.4)
*
1: The upper frequency is limited by the specifications of the main frame in which this unit is installed.
*
2: Applicable when this unit is installed in MG3670B/C, MG3671A/B and MG3672A. Not applicable when this unit is installed in MG3670A.