E4418B/9B EPM Series power meter performance characteristics

(continued)

Table 1

Table 2

No. of averages

Noise multiplier

Normal mode

x2 mode

1

5.5

6.5

2

3.89

4.6

4

2.75

3.25

8

1.94

2.3

16

1

1.63

32

0.85

1

64

0.61

0.72

128

0.49

0.57

256

0.34

0.41

512

0.24

0.29

1024

0.17

0.2

Model

Zero set

±500 pW

±500 nW

±5 nW

±50 pW

±25 nW

±25 μW

±2.5 μW

±50 nW

±50 μW

±20 pW

±5 μW

±30 pW

±200 nW

Zero drift

1

<±150 pW

<±150 nW

<±1.5 nW

<±15 pW

<±3 nW

<±3 μW

<±0.3 μW

<±10 nW

<±10 μW

<±4 pW

<±1 μW

<±6 pW

<±40 nW

Measurement noise

2

<700 pW

<700 nW

<7 nW

<70 pW

<80 nW

<80 μW

<8 μW

<110 nW

<110 μW

<45 pW

<10 μW

<65 pW

<450 nW

1. Within one hour after zero set, at a constant temperature, after
a 24-hour warm-up of the power meter.
2. The number of averages at 16 for normal mode and 32 for x2
mode, at a constant temperature, measured over a one-minute
interval and two standard deviations. For E-Series sensors, the
measurement noise is measured within the low range. Refer to
the relevant sensor manual for further information.
3. Specification applies to the low power path, 15% to 75% relative
humidity.

E9300A, E9301A, E9304A

3

E9300B, E9301B

3

E9300H, E9301H

3

E4412A, E4413A

N8481A, N8482A, N8485A,N8487A,
N8486AQ, N8486AR

N8481B, N8482B

N8481H, N8482H

8481A, 8482A, 8483A, 8485A, 8487A,
R8486A, Q8486A

8481B, 8482B

8481D, 8485D, 8487D

8481H, 8482H

R8486D, Q8486D

V8486A, W8486A

5

Settling time

1

Table 3

No. of averages

Settling time with
E-Series sensors (s)

Normal mode

x2 mode

Settling time with N8480
Series sensors (s)

Normal mode

x2 mode

Settling time with 8480
Series sensors (s)

Normal mode

x2 mode

1

0.07

0.04

0.15

0.15

0.15

0.15

2

0.12

0.07

0.2

0.18

0.2

0.18

4

0.21

0.12

0.3

0.22

0.3

0.22

8

0.4

0.21

0.5

0.35

0.5

0.35

16

1

0.4

1.1

0.55

1.1

0.55

32

1.8

1

1.9

1.1

0.9

1.1

64

3.3

1.8

3.4

1.9

3.4

1.9

128

6.5

3.4

6.6

3.5

6.6

3.5

256

13

6.8

13

6.9

13

6.9

512

27

14.2

27

14.5

27

14.5

1024

57

32

57

33

57

33

Manual filter, 10-dB decreasing power step for normal and x2 modes (not across range switch points for
E-Series and N8480 Series sensors)

*E-Series sensors in Fast mode (using free run trigger), within the range –50 dBm to +17 dBm, the settling time is:

E4418B: 10 ms

2

E4419B: 20 ms

2

1. Settling time: 0 to 99% settled readings over the GPIB
2. When a power step crosses through the sensor’s auto-range switch point, add 25 ms. Refer to the relevant sensor manual for switch point information.