E-Series CW power sensor specifications

10

Widest dynamic range: 100 pW to 100 mW (–70 dBm to +20 dBm)

Calibration factor (CF) and reflection coefficient (Rho)

Table 4

Model

Maximum SWR

10 MHz – 18 GHz

50 MHz – 26.5 GHz

Maximum SWR

*10 MHz to <30 MHz: 1.22
30 MHz to <2 GHz: 1.15
2 GHz to <6 GHz: 1.17
6 GHz to <11 GHz:1.2
11 GHz to <18 GHz: 1.27

50 MHz to <100 MHz: 1.21
100 MHz to <8 GHz: 1.19
8 GHz to <18 GHz: 1.21
18 GHz to 26.5 GHz: 1.26

Maximum power

200 mW (+23 dBm)

200 mW (+23 dBm)

Connector type

Type-N (m)

APC-3.5 mm (m)

Calibration factor and reflection
coefficient data are provided at 1 GHz
increments on a data sheet included
with the power sensor. This data is
unique to each sensor. If you have more
than one sensor, match the serial number
on the Certificate of Calibration (COC)
with the serial number on the power
sensor you are using. The CF corrects
for the frequency response of the sensor.
The EPM power meter automatically
reads the CF data stored in the sensor
and uses it to make the corrections.
For power levels greater than 0 dBm,
add 0.5%/dB to the calibration factor
uncertainty specification.

Reflection coefficient (Rho) relates
to the SWR according to the following
formula: SWR = 1 + Rho/1 - Rho

Maximum uncertainties of the CF data
are listed in Table 5a, for the E4412A
power sensor, and Table 5b for the
E4413A power sensor. The uncertainty
analysis for the calibration of the sensors
was done in accordance with the
ISO/TAG4 Guide. The uncertainty data
reported on the calibration certificate is
the expanded uncertainty with a 95%
confidence level and a coverage
factor of 2.

Table 5a. E4412A calibration factor
uncertainty at 1mW (0 dBm)

Frequency

10 MHz

30 MHz

50 MHz

100 MHz

1.0 GHz

2.0 GHz

4.0 GHz

6.0 GHz

8.0 GHz

10.0 GHz

11.0 GHz

12.0 GHz

14.0 GHz

16.0 GHz

18.0 GHz

Table 5b. E4413A calibration factor
uncertainty at 1mW (0 dBm)

Frequency

50 MHz

100 MHz

1.0 GHz

2.0 GHz

4.0 GHz

6.0 GHz

8.0 GHz

10.0 GHz

11.0 GHz

12.0 GHz

14.0 GHz

16.0 GHz

17.0 GHz

18.0 GHz

20.0 GHz

24.0 GHz

26.0 GHz

28.0 GHz

Uncertainty*(%)

1.8

1.8

Reference

1.8

1.8

2.4

2.4

2.4

2.4

2.4

2.4

2.4

2.4

2.6

2.6

Uncertainty*(%)

Reference

1.8

1.8

2.4

2.4

2.4

2.4

2.6

2.6

2.8

2.8

2.8

2.8

2.8

3.0

3.0

3.0

3.0

* Applies to sensors with serial prefix US 3848 or greater

E4412A

E4413A