E-Series average power sensor specifications
The E-Series E9300 power sensors have
two independent measurement path as
shown:
Table 7
Table 8
Model
E9300A
E9301A
E9304A
E9300B
E9301B
E9300H
E9301H
Frequency range
10 MHz to 18 GHz
10 MHz to 6 GHz
9 kHz to 6 GHz
10 MHz to 18 GHz
10 MHz to 6 GHz
10 MHz to 18 GHz
10 MHz to 6 GHz
–60 dBm to +20 dBm wide dynamic range sensors
–30 dBm to +44 dBm wide dynamic range sensors
–50 dBm to +30 dBm wide dynamic range sensors
Maximum SWR
(25 °C ± 10 °C)
10 MHz to 30 MHz: 1.15
30 MHz to 2 GHz: 1.13
2 GHz to 14 GHz: 1.19
14 GHz to 16 GHz: 1.22
16 GHz to 18 GHz: 1.26
10 MHz to 30 GHz: 1.15
30 MHz to 2 GHz: 1.13
2 GHz to 6 GHz: 1.19
9 kHz to 2 GHz: 1.13
2 GHz to 6 GHz: 1.19
10 MHz to 8 GHz: 1.12
8 to 12.4 GHz: 1.17
12.4 to 18 GHz: 1.24
10 MHz to 6 GHz: 1.12
10 MHz to 8 GHz: 1.15
8 to 12.4 GHz: 1.25
12.4 to 18 GHz: 1.28
10 MHz to 6 GHz: 1.15
Maximum SWR
(0 to 55 °C)
10 MHz to 30 MHz: 1.21
30 MHz to 2 GHz: 1.15
2 GHz to 14 GHz: 1.20
14 GHz to 16 GHz: 1.23
16 GHz to 18 GHz: 1.27
10 MHz to 30 MHz: 1.21
30 MHz to 2 GHz: 1.15
2 GHz to 6 GHz: 1.20
9 kHz to 2 GHz: 1.15
2 GHz to 6 GHz: 1.20
10 MHz to 8 GHz: 1.14
8 to 12.4 GHz: 1.18
12.4 to 18 GHz: 1.25
10 MHz to 6 GHz: 1.14
10 MHz to 8 GHz: 1.17
8 to 12.4 GHz: 1.26
12.4 to 18 GHz: 1.29
10 MHz to 6 GHz: 1.17
Maximum power
+25 dBm (320 mW) average;
+33 dBm peak (2 W)
(< 10 μsec)
+25 dBm (320 mW) average;
+33 dBm peak (2 W)
(< 10 μsec)
+25 dBm (320 mW) average;
+33 dBm peak (2 W)(< 10 μsec)
0 to 35 °C: 30 W avg
35 to 55 °C: 25 W avg
< 6 GHz: 500 W pk
> 6 GHz: 125 W pk
500 W.μS per pulse
0 to 35 °C: 30 W avg
35 to 55 °C: 25 W avg
< 6 GHz: 500 W pk
> 6 GHz: 125 W pk
500 W.μS per pulse
3.16 W avg
100 W pk
100 W.μS per pulse
3.16 W avg
100 W pk
100 W.μS per pulse
Connector type
Type-N (m)
Type-N (m)
Type-N (m)
Type-N (m)
Type-N (m)
Type-N (m)
Type-N (m)
“A” suffix sensors
–10 to +20 dBm
–60 to –10 dBm
“B” suffix sensors
+20 to +44 dBm
–30 to +20 dBm
“H” suffix sensors
0 to +30 dBm
–50 to 0 dBm
High power path
Low power path
Specifications are valid ONLY after proper
calibration of the power meter and apply
for CW signals unless otherwise stated.
Specifications apply over the temperature
range 0 °C to 55 °C unless otherwise
stated, and specifications quoted over
the temperature range 25 °C ±10 °C,
conform to the standard environmental
test conditions as defined in TIA/EIA/
IS-97-A and TIA/EIA/IS-98-A [1].