Power Sensor Specifications
ML2430A, ML2480B, ML2490A
Power Sensor Performance
Sensor
Frequency
Range
CW Dynamic Range
(dBm)
SWR
Rise Time
1
(ms)
Sensor Linearity
7
RF
Connector
2
Standard Diode Sensors
MA2472D 10 MHz to 18 GHz
–70 to +20
CW mode
–43 to +20
(ML243xA, Profile mode)
–37 to +20
(ML2480A/B or ML2490A,
Pulse/Mod mode)
< 1.17; 10 MHz to 150 MHz
< 1.12; 150 MHz to 2 GHz
< 1.22; 2 GHz to 12.4 GHz
< 1.25; 12.4 GHz to 18 GHz
< 1.35; 18 GHz to 32 GHz
< 1.50; 32 GHz to 40 GHz
< 1.63; 40 GHz to 50 GHz
< 0.004
< 1.8 %, ≤18 GHz
< 2.5 %, ≤40 GHz
< 3.5 %, ≤50 GHz
for MA2475D
(see Note 4)
N(m)
MA2473D 10 MHz to 32 GHz
K(m)
MA2474D 10 MHz to 40 GHz
K(m)
MA2475D 10 MHz to 50 GHz
V(m)
Temperature accuracy: < 1 % < 40 GHz, < 1.5 % < 50 GHz, 5 ºC to 50 ºC
High Accuracy Diode Sensors
MA2442D 10 MHz to 18 GHz
–67 to +20 CW mode
–40 to +20
(ML243xA, Profile mode)
–34 to +20
(ML2480A/B or ML2490A,
Pulse/Mod mode)
< 1.17; 10 MHz to 150 MHz
< 1.08; 150 MHz to 2 GHz
< 1.16; 2 GHz to 12.4 GHz
< 1.21; 12.4 GHz to 18 GHz
< 1.29; 18 GHz to 32 GHz
< 1.44; 32 GHz to 40 GHz
< 1.50; 40 GHz to 50 GHz
< 0.004
< 1.8 %, ≤18 GHz
< 2.5 %, ≤40 GHz
< 3.5 %, ≤50 GHz
for MA2445D
(see Note 5)
N(m)
MA2444D 10 MHz to 40 GHz
K(m)
MA2445D 10 MHz to 50 GHz
V(m)
Temperature accuracy: < 1 % < 40 GHz, < 1.5 % < 50 GHz, 5 ºC to 50 ºC
Universal Power Sensors
MA2481D
10 MHz to 6 GHz
–60 to +20
< 1.17; 10 MHz to 150 MHz
< 1.12; 150 MHz to 2 GHz
< 1.22; 2 GHz to 6GHz
< 1.22; 6 GHz to 12.4 GHz
< 1.25; 12.4 GHz to 18 GHz
< 0.004
with option
1 only
< 3 %, ≤6 GHz
< 3 %, ≤18 GHz
(1.8 % CW
with option 1)
N(m)
MA2482D 10 MHz to 18 GHz
Option 01 Adds fast CW mode to Universal Power Sensors for high speed measurements of CW signal plus TDMA and pulse measurements.
Temperature accuracy: < 1 %, 15 ºC to 35 ºC
Wideband Sensors
MA2490A
3
50 MHz to 8 GHz
CW Mode:
–60 to +20
Pulse/Modulated Mode:
–30 to +20
(with ML2480B/90A)
< 1.17; 50 MHz to 150 MHz
< 1.12; 150 MHz to 2.5 GHz
< 1.22; 2.5 GHz to 8 GHz
< 1.22; 8 GHz to 12.4 GHz
< 1.25; 12.4 GHz to 18 GHz
< 18 ns
< 7 %
50 MHz to 300 MHz
< 3.5 %
300 MHz to 8 GHz
N(m)
MA2491A
3
50 MHz to 18 GHz
Temperature accuracy: < 1 % 10 ºC to 45 ºC
Pulse Sensor
MA2411B
300 MHz to
40 GHz
–20 to +20 dBm
< 1.15; 300 MHz to 2.5 GHz
< 1.35; 2.5 GHz to 26 GHz
< 1.50; 26 GHz to 40 GHz
< 8 ns
typical
12 ns
maximum
< 18 ns
when used
with
ML2487B/M
L2488B
< 4.5 %
300 MHz to 18 GHz
< 7 %
18 GHz to 40 GHz
K(m)
Requires 1 GHz Calibrator (Option 15) to be fitted on the meter, if used with ML248xB.
Temperature accuracy: < 2 % 10 ºC to 45 ºC
Thermal Sensor
MA24002A 10 MHz to 18 GHz
–30 dBm to +20 dBm
< 1.90; 10 MHz to 50 MHz
< 1.17; 50 MHz to 150 MHz
< 1.10; 150 MHz to 2 GHz
< 1.15; 2 GHz to 12.4 GHz
< 1.20; 12.4 GHz to 18 GHz
< 1.25; 18 GHz to 32 GHz
< 1.30; 32 GHz to 40 GHz
< 1.40; 40 GHz to 50 GHz
< 15
1.8 % < 18 GHz
2.0 % < 40 GHz
2.5 % < 50 GHz
(see note 6)
N(m)
MA24004A 10 MHz to 40 GHz
K(m)
MA24005A 10 MHz to 50 GHz
V(m)
Temperature accuracy: < 1 % < 30 GHz < +10 dBm, < 1.5 % ≥30 GHz ≥ +10 dBm
1. 0.0 dBm, room temperature with standard 1.5m sensor cable.
2. Each MA2400A/D Series sensor incorporates precision RF connectors with
hexagon coupling nut for attachment by industry standard torque wrench.
3. MA2490/1A and MA2411B sensors must be used with ML2480B or ML2490A
series power meters.
4. MA2475D Linearity applicable from –70 to +15 dBm.
Add 1 % for power levels > +15 dBm
2000-1537-R supplied as standard with the power meter.
5. MA2445D Linearity applicable from –67 to +15 dBm.
Add 1 % for power levels > +15 dBm
6. MA24005D Linearity applicable from –30 to +15 dBm.
Add 1 % for power levels > +15 dBm
7. Sensor linearity specifications are ± value.
Pulse/modulated performance only specified with 1.5m sensor cable length