Reference calibrator, General specifications – Atec Anritsu-ML2400 Series User Manual
Page 7

ML2430A, ML2480B, ML2490A
Power Meter Specifications
ML2430A, ML2480B, ML2490A TDS
PN: 11410-00423 Rev. H
7
Reference Calibrator
General Specifications
ML2430A Series
ML2480B Series
ML2490A Series
Reference Calibrator Power 1 mW
Power Accuracy (Traceable to
National Standards) ± 1.2 % per year
Frequency 50 MHz (nominal)
50 MHz (standard)
1 GHz (optional)
50 MHz and 1 GHz (both standard)
Frequency Accuracy < 1 %
< 1 % (50 MHz)
< 2 % (1 GHz)
VSWR < 1.04
< 1.12 (50 MHz)
< 1.2 (1 GHz)
Connector Type N female
ML2430A Series
ML2480B Series
ML2490A Series
General MIL-T28800F, class 3
Non Volatile
RAM Battery
Lithium (10 year life)
Lithium (5 year life)
Battery Option > 6 hr usable with 3000 mAhr
(NiMH) battery
N/A
DC Power Requirements 12 to 24 VDC, Reverse protected to
–40 V Maximum input 30 V
N/A
AC Power Requirements
85 VAC to 264 VAC
47 Hz to 440 Hz
40 VA Maximum
85 VAC to 264 VAC
47 Hz to 440 Hz
EMI, EMC, Safety Complies with requirements for CE marking EN 61326, EN61010-1
Operating Temperature 0 °C to 50 °C
Storage Temperature –40 °C to 70 °C
Moisture Splash and rain resistant, 95 % humidity non-condensing
Dimensions 223 mm x 150 mm x 390 mm
Weight 3 kg (excluding battery option)
3 kg
Warranty 1 year Standard, 3 year Optional