Operation, Features – Atec Anritsu-ML2400 Series User Manual
Page 4
4
PN: 11410-00423 Rev. H
ML2430A, ML2480B, ML2490A TDS
Power Meter Specifications
ML2430A, ML2480B, ML2490A
Operation
Features
(summary)
ML2430A Series
ML2480B Series
ML2490A Series
Measurement Display
Readout (Numerical) 2
2 (CW or Pulse/Modulated measurement modes)
Measurement Display Profile
(Graph)
Power vs. Time graphic of readout
data or
Profile of Peak power for analysis of
repetitive pulse or transient
waveforms
2 (Pulse/Modulated measurement mode)
Source sweep Single channel power sweep or frequency sweep
Peaking meter ± 5 dB range CW (Readout mode) only
Amplifier Range
Dynamic range covered by five
overlapping amplifier ranges:
R1, R2, R3, R4, and R5.
Universal Sensor MA2481/82D
ranges 1 to 6.
Pulse modulated mode:
Dynamic range covered by three overlapping amplifier ranges:
R7, R8, and R9.
CW mode:
Dynamic range covered by five overlapping amplifier ranges:
R1, R2, R3, R4, and R5.
Universal Sensor MA2481/82D ranges 1 through 6
Range Hold
Auto or Manual
(current range or selectable 1
through 5).
Automatic or manual. When in manual, clear indication is given to user
(display and GPIB) of fault conditions (under or over range).
ML2430A Series
ML2480B Series
ML2490A Series
Display Monochrome LCD, with backlight
and adjust- able contrast
Color LCD
Display resolution in
Readout mode
0.1 dB to 0.001 dB
Linear power units, 3 to 6 digits,
1 to 3 digits selectable to right of
decimal, nW to W
Voltage, 1 to 2 digits selectable to
right of decimal
0.1 dB to 0.001 dB
Display resolution in
Profile mode
0.01 dB
Time measurement
resolution
Profile and P vs. T modes:
200 pixels display resolution
For a 1 ms Profile window, cursor
resolution on the display is 5 µs
16 ns Pulse/Modulated mode
15 µs CW Mode
1 ns (RRS mode)
16 ns (non RRS mode)
Pulse/Modulated mode
15 µs CW Mode
Measurement hold Hold, Max, Min
Measurements Average, Min, Max
Average, Min, Max, Peak, Crest, PAE (Power Added Efficiency)
Power statistics —
PDF, CDF, CCDF
Voltage measurement range 0.00 to 20.00 V nominal
Display units (Lin) Display
units (Log)
Watt, %, VoltsdBm, dB, dBµV,
dBmV, dBr
dBm, dBW, dB, dBµV, dBmV
Display range –199.99 dB to +199.99 dB
Measurement Gates 1
Four Independently set Gates or eight repeated Gates
One Fence per Measurement gate
Gate measurement supports Average, Peak, Crest, Max and Min
Markers 2
Four Markers and One Delta Marker, Marker to Max/Min, Pulse
Rise/Fall-time, Pulse Width, Off Period, Pulse Repetition Interval
Rise Fall/Search Parameter Variable % Reference: Max Marker or Gate
Power Level
Limit lines
Fixed value high and low limits with
audible, rear panel TTL output,
and/or visible Pass/Fail alarm
indication
Failure indication can latch for
transient failure detection
Simple pass/fail for CW
Complex limits for pulsed and TDMA systems
30 Limits Stores available on the instrument
Offset range –199.99 dB to +199.99 dB (Fixed value or frequency dependent table)