beautypg.com

Averaging, Triggering – Atec Anritsu-ML2400 Series User Manual

Page 5

background image

ML2430A, ML2480B, ML2490A

Power Meter Specifications

ML2430A, ML2480B, ML2490A TDS

PN: 11410-00423 Rev. H

5

Averaging

Triggering

ML2430A Series

ML2480B Series

ML2490A Series

Type Auto (Moving), Manual (Moving, Repeat)

Range 1 to 512

Low-level Averaging

Low, Medium and High settings

apply post average low pass filter

to improve visibility at high display

resolution.

N/A

ML2430A Series

ML2480B Series

ML2490A Series

Source

Internal, External (TTL or RF

Blanking), GPIB, Manual, or

Continuous.

Continuous (not in Random Repetitive Sampling mode) Internal, External

TTL (Rising or falling Edge), GPIB, or external Bus.

Trigger Modes

Manual

Single power value set to cover entire measurement dynamic range of sensor.
Auto

Automatically sets trigger level for signal over measurement dynamic range.

Nominal Internal Trigger

Bandwidth

N/A

Variable-auto set and manual
20 MHz, 2 MHz, 200 kHz, 20 kHz

Arming Sources

Sets the trigger arming, unless the

trigger source is set to EXTTTL
When ARMING is set to Blanking

ON, only samples taken when the

rear panel Digital Input BNC is

active will be averaged in the

measurement.

Repetitive Sampling Modes:

Automatic

Frame for QAM and multi-pulse
Continuous Sampling Modes:

Single

Automatic

Frame for QAM and multi-pulse

Frame Arming

Time Range

N/A

0 to 64 x trigger capture time range or 120 s, whichever is greater.

Internal Trigger Dynamic

Range

–15 dBm to +20 dBm (all diode

sensors, selectable to –25 dBm)

–28 dBm to +10 dBm with MA2472D CW mode
–18 dBm to +14 dBm with MA2491A
–30 dBm to +10 dBm with MA2472D Pulse/Modulated mode

Internal Trigger

Level Accuracy (typical) 1 dB

Internal Trigger

Settable Resolution 0.1 dB

Trigger Time

Resolution Uncertainty N/A

±2 ns or display resolution, whichever is larger.

(Trigger Capture time 50 ns to 3.2 µs)
±16 ns or display resolution whichever is larger.

(Trigger Capture time 3.2 µs to 7 s)

Trigger Delay Range 0 ms to 999 ms

Pulse modulated mode:

Pretrigger (-ve): 95 % of the Trigger Capture range

Post Trigger: Set by 256K buffer and sample rate
CW mode:

Post Trigger Only: 0 ms to 999 ms depending on Trigger Capture period

setting.

External Trigger Range TTL rising or falling edge (BNC input)

Pre-trigger Range N/A

90 % of trigger capture range

Trigger Delay

Settable Resolution 0.5 % of display period or 100 ns

200 display points

1 ns or 0.5 % of trigger capture time, whichever is larger.
400 display points

1 ns or 0.25 % of trigger capture time (400 points), whichever is larger.

Trigger Delay Uncertainty N/A

± 2 ns for pre and post trigger

(Trigger capture time of 3.2 µs or 50 ns)

Trigger Latency N/A

± 15 ns (20 MHz trigger BW)

Trigger/Display

Capture Range

Profile mode: 10 ms to 7 s
P v T mode: 1 m to 24 hrs

3.2 µs to 7 s

50 ns to 7 s

Trigger Capture Time

Settable Resolution N/A

200 display points

16 ns or 0.5 % of trigger capture

time, whichever is larger
400 display Points

16 ns or 0.25 % of trigger capture

time, whichever is larger.

200 display points

1 ns or 0.5 % of trigger capture

time, whichever is larger
400 display Points

1 ns or 0.25 % of trigger capture

time, whichever is larger.

Trigger Point Display

(on-screen)

On-screen indicator/message

Trigger point depicted by trigger edge waveform (edge represents trigger

point of signal). Display position of trigger edge waveform adjustable.