Performance characteristics, Continued) – Atec Agilent-5000 Series User Manual
Page 19
![background image](/manuals/318400/19/background.png)
Performance characteristics
(continued)
General characteristics
Physical size
35.4 cm wide x 18.8 cm high x 17.4 cm deep (without handle)
38.5 cm wide x 18.8 cm high x 17.4 cm deep (with handle)
Weight
Net: 4.1 kg (9 lbs)
Shipping: approximately 9 kgs (20 lbs)
Probe comp output frequency ~ 1.2 kHz, amplitude ~2.5 V
Trigger out
0 to 5 V into open circuit (~23 ns delay)
0 to 2.5 V into 50 Ω
Kensington lock
Connection on rear panel for security
Power requirements
Line voltage range
100-120 V, 50/60/400 Hz; 100-240V, 50/60 Hz auto ranging
Line frequency
50/60 Hz, 100-240 VAC; 440 Hz, 100-132 VAC
Power usage
120 W max
Environmental characteristics
Ambient temperature
Operating -10 °C to +55 °C; non-operating –40 °C to +70 °C
Humidity operating
95% RH at 40 °C for 24 hour; non-operating 90% RH at 65 °C for 24 hour
Altitude
Operating to 4,570 m (15,000 ft); non-operating to 15,244 m (50,000 ft)
Vibration
Agilent class GP and MIL-PRF-28800F; Class 3 random
Shock
Agilent class GP and MIL-PRF-28800F (operating 30 g, 1/2 sine, 11-ms duration, 3 shocks/axis along
major axis. Total of 18 shocks)
Pollution degree
Normally only dry non-conductive pollution occurs. Occasionally a temporary conductivity caused by
condensation must be expected.
Indoor use
Rated for indoor use only
Other
Measurement categories
CAT I
Regulatory information
Safety IEC 61010-1:2001 / EN 61010-1:2001
Canada: CSA C22.2 No. 61010-1:2004
USA: UL 61010-1:2004
Supplementary information
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the
EMC Directive 89/336/EEC, and carries the CE-marking accordingly.
The product was tested in a typical configuration with HP/Agilent test systems.
Product specifications, characteristics, and descriptions in this document are subject to change
without notice.
19