Performance characteristics

(continued)

Horizontal

Range

DSO501xA: 5 nsec/div to 50 sec/div

DSO503xA: 2 nsec/div to 50 sec/div

DSO505xA: 1 nsec/div to 50 sec/div

Resolution

2.5 psec

Time scale accuracy*

Specification: 25 ppm from 0 °C to 40 °C

Characteristic: 25 ppm + 1 ppm per °C from 40 °C to 55 °C

Vernier

1-2-5 increments when off, ~25 minor increments between major settings when on

Delay range

Pre-trigger (negative delay): Greater of 1 screen width or 125 µs

Post-trigger (positive delay): 1 s to 500 seconds

Channel delta-t accuracy

Same channel: ±0.0025% reading ±0.1% screen width ±20 ps

Channel-to-channel: ±0.0025% reading ±0.1% screen width ±40 ps

Same channel example (DSO505xA): For signal with pulse width of 10 µs, oscilloscope set to 5 µs/div

(50 µs screen width), delta-t accuracy = ±{0.0025% (10 µs) + 0.1% (50 µs) + 20 ps} = 50.27 ns

Modes

Main, delayed, roll, XY

XY

Bandwidth: Max bandwidth

Phase error at 1 MHz: < 0.5 degrees

Z Blanking: 1.4 V blanks trace (use external trigger on DSO50x2A, channel 4 on DSO50x4A)

Reference positions

Left, center, right

Segmented memory re-arm time 8 µs (minimum time between trigger events)

Trigger system

Sources

DSO50x2A: Ch 1, 2, line, ext

DSO50x4A: Ch 1, 2, 3, 4, line, ext

Modes

Auto, normal (triggered), single

Holdoff time

~60 ns to 10 seconds

Trigger jitter

15 ps rms

* Denotes warranted specification. Specifications are valid after a 30 minute warm-up period and ±10 °C from firmware calibration procedure.

15