Performance characteristics

(continued)

Channel triggering

Range (internal)

±6 div from center screen

Sensitivity

1

< 10 mV/div: greater of 1 div or 5mV; ≥ 10 mV/div: 0.6 div

Coupling

AC (~10 Hz), DC, noise reject, HF reject and LF reject (~50 kHz)

1 Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10° C from firmware calibration temperature.

External (EXT) triggering

DSO50x2A

DSO50x4A

Input impedance

1 MΩ ±3% || 12 pF or 50 Ω ±1%

1.015 kΩ ±5%

Maximum input

CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk

±15 V

CAT II 100 Vrms, 400 Vpk

with N2863A 10:1 probe: CAT I 600 V, CAT II 300 V (DC + peak AC)

with 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk

5 Vrms with 50-Ω input, CAT I

Range

DC coupling: trigger level ± 1 V and ± 8 V

±5 V

Sensitivity

For ± 1V range setting:

DC to 100 MHz, 500 mV

DC to 100 MHz, 100 mV, >100 MHz to bandwidth of

the oscilloscope, 200 mV

For ±8 V range setting:

DC to 100 MHz, 250 mV; >100 MHz to bandwidth of

the oscilloscope, 500 mV

Coupling

AC (~10 Hz), DC, noise reject, HF reject and LF reject (~50 kHz)

Probe ID

Auto probe sense and AutoProbe interface

Agilent- and Tektronix-compatible passive probe sense

Display system

Display

6.3‑inch (161 mm) diagonal color TFT LCD

Display update rate

Up to 100,000 waveforms/sec in real-time mode

Resolution

XGA:

768 vertical by 1024 horizontal points (screen area);

640 vertical by 1000 horizontal points (waveform area)

256 levels of intensity scale

Controls

Waveform intensity on front panel. Vectors on/off; infinite persistence on/off,

8 x 10 grid with intensity control

Built-in help system

Key-specific help displayed by pressing and holding key or softkey of interest

Real-time clock

Time and date (user adjustable)