Performance characteristics

Acquisition

Sample rate

DSO501xA/503xA: 2 GSa/a each channel

DSO505xA: 4 GSa/s half channel

1

, 2 GSa/s each channel

Equivalent-time sample rate

400 GSa/s (when real-time mode is turned off)

Memory depth

Standard 8 Mpts half channel1 , 4 Mpts each channel

Vertical resolution

8 bits, up to 12 in high-resolution or averaging modes

Peak detection

DSO501xA: 1 ns peak detect

DSO503xA: 500 ps peak detect

DSO505xA: 250 ps peak detect

Averaging

Selectable from 2, 4, 8, 16, 32, 64 … to 65536

High resolution mode

Average mode with #avg = 1

12 bits of resolution when ≥10 µs/div, at 4 GSa/s or ≥20 µs/div, at 2 GSa/s

Filter

Sinx/x interpolation (single shot BW = sample rate/4 or bandwidth of oscilloscope, whichever is less)

with vectors on and in real-time mode

Vertical system

Scope channels

DSO50x2A: 2 Ch 1 and 2 simultaneous acquisition

DSO50x4A: Ch 1, 2, 3, and 4 simultaneous acquisition

Bandwidth (–3dB)

2

DSO501xA: DC to 100 MHz

DSO503xA: DC to 300 MHz

DSO505xA: DC to 500 MHz

AC coupled

DSO501xA: 3.5 Hz to 100 MHz

DSO503xA: 3.5 Hz to 300 MHz

DSO505xA: 3.5 Hz to 500 MHz

Calculated rise time

DSO501xA: 3.5 nsec

(= 0.35/bandwidth)

DSO503xA: 1.17 nsec

DSO505xA: 700 psec

Single-shot bandwidth

DSO501xA: 100 MHz

DSO503xA: 300 MHz

DSO505xA: 500 MHz

1 Half channel is when only one channel of channel pair 1-2 is turned on, or one channel of channel pair 3-4 is turned on.

2 Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10° C from firmware calibration temperature.

13