Performance characteristics – Atec Agilent-5000 Series User Manual
Page 13
Performance characteristics
Acquisition
Sample rate
DSO501xA/503xA: 2 GSa/a each channel
DSO505xA: 4 GSa/s half channel
1
, 2 GSa/s each channel
Equivalent-time sample rate
400 GSa/s (when real-time mode is turned off)
Memory depth
Standard 8 Mpts half channel1 , 4 Mpts each channel
Vertical resolution
8 bits, up to 12 in high-resolution or averaging modes
Peak detection
DSO501xA: 1 ns peak detect
DSO503xA: 500 ps peak detect
DSO505xA: 250 ps peak detect
Averaging
Selectable from 2, 4, 8, 16, 32, 64 … to 65536
High resolution mode
Average mode with #avg = 1
12 bits of resolution when ≥10 µs/div, at 4 GSa/s or ≥20 µs/div, at 2 GSa/s
Filter
Sinx/x interpolation (single shot BW = sample rate/4 or bandwidth of oscilloscope, whichever is less)
with vectors on and in real-time mode
Vertical system
Scope channels
DSO50x2A: 2 Ch 1 and 2 simultaneous acquisition
DSO50x4A: Ch 1, 2, 3, and 4 simultaneous acquisition
Bandwidth (–3dB)
2
DSO501xA: DC to 100 MHz
DSO503xA: DC to 300 MHz
DSO505xA: DC to 500 MHz
AC coupled
DSO501xA: 3.5 Hz to 100 MHz
DSO503xA: 3.5 Hz to 300 MHz
DSO505xA: 3.5 Hz to 500 MHz
Calculated rise time
DSO501xA: 3.5 nsec
(= 0.35/bandwidth)
DSO503xA: 1.17 nsec
DSO505xA: 700 psec
Single-shot bandwidth
DSO501xA: 100 MHz
DSO503xA: 300 MHz
DSO505xA: 500 MHz
1 Half channel is when only one channel of channel pair 1-2 is turned on, or one channel of channel pair 3-4 is turned on.
2 Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10° C from firmware calibration temperature.