Performance characteristics

(continued)

Vertical system (continued)

Range

1

2 mV/div to 5 V/div (1 MΩ or 50 Ω)

Maximum input

Maximum input voltage for analog inputs:

CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk

CAT II 100 Vrms, 400 Vpk

with N2863A 10:1 probe: CAT I 600 V, CAT II 300 V (DC + peak AC)

with 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk

with 50 Ω input: 5 Vrms, CAT I

Offset range

±5 V on ranges < 10 mV/div;

±20 V on ranges 10 mV/div to 200 mV/div;

±75 V on ranges >200 mV/div

Dynamic range

±8 div

Input impedance

1 MΩ ± 1% || 12 pF or 50 Ω ± 1.0%, selectable

Coupling

AC, DC

BW limit

25 MHz selectable

Channel-to-channel isolation

DC to max bandwidth > 40 dB

Standard probes

DSO501xA: 10:1 N2863A shipped standard for each oscilloscope channel

DSO503xA: 10:1 N2863A shipped standard for each oscilloscope channel

DSO505xA: 10:1 10073C shipped standard for each oscilloscope channel

Probe ID

Auto probe sense and AutoProbe interface

Agilent- and Tektronix-compatible passive probe sense

ESD tolerance

±2 kV

Noise, RMS, input shorted

DSO501xA: 0.5% full scale or 250 µV, whichever is greater

DSO503xA: 0.5% full scale or 300 µV, whichever is greater

DSO505xA: 0.5% full scale or 360 µV, whichever is greater

DC vertical gain accuracy

2

±2.0% full scale

DC vertical offset accuracy

≤ 200 mV/div: ±0.1 div ±2.0 mV ±0.5% offset value;

> 200 mV/div: ±0.1 div ±2.0 mV ±1.5% offset value

Single cursor accuracy

1

±{DC vertical gain accuracy + DC vertical offset accuracy + 0.2% full scale (~1/2 LSB)}

Example: for 50 mV signal, oscilloscope set to 10 mV/div (80 mV full scale), 5 mV offset,

accuracy = ±{2.0% (80 mV) + 0.1 (10 mV) + 2.0 mV + 0.5% (5 mV) + 0.2%(80 mV)} = ± 4.785 mV

Dual cursor accuracy

1

±{DC vertical gain accuracy + 0.4% full scale (~1 LSB)}

Example: for 50 mV signal, oscilloscope set to 10 mV/div (80 mV full scale), 5 mV offset,

accuracy = ±{2.0% (80 mV) + 0.4% (80 mV)} = ±1.92 mV

1 2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 2 mV/div sensitivity setting.

2 Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.