14
Infiniium 54850 Series Performance Characteristics
(continued)
Computer system and peripherals, I/O ports (continued)
I/O ports
LAN
RJ-45 connector, supports 10Base-T and 100Base-T. Enables Web-enabled remote
control, e-mail on trigger or demand, data/file transfers and network printing.
GPIB
IEEE 488.2, fully programmable.
RS-232 (serial)
COM1, printer and pointing device support.
Parallel
Centronics printer port.
PS/2
2 ports. Supports PS/2 pointing and input devices.
USB
2 ports. Allows connection of USB peripherals and pointing devices while the
oscilloscope is on.
Video output
15 pin VGA, full color output of scope waveform display.
Dual-monitor video output
15 pin XGA, full color output for using third-party applications.
Auxiliary output
DC (±2.4 V); square wave (~715 Hz and 456 MHz); trigger output
(255 mV p-p into 50
Ω
).
Trigger output
5 V 50
Ω
back-terminated.
Time base reference output
10 MHz, 5V 50
Ω
back-terminated.
General characteristics
Temperature
Operating: 5° C to +40° C.
Non-operating: –40° C to +70° C.
Humidity
Operating: Up to 95% relative humidity (non-condensing) at +40°C.
Non-operating: Up to 90% relative humidity at +65°C.
Altitude
Operating: Up to 4,600 meters (15,000 feet).
Non-operating: Up to 15,300 meters (50,000 feet).
Vibration
Operating: Random vibration 5-500 Hz, 10 minutes per axis, 0.3 g(rms).
Non-operating: Random vibration 5-500 Hz, 10 minutes per axis, 2.41 g(rms);
resonant search 5-500 Hz, swept sine, 1 octave/minute sweep rate, (0.75g),
5 minute resonant dwell at 4 resonances per axis.
Power
100-240 VAC, ± 10%, Cat II, 47 to 440 Hz; max power dissipated: 390 W.
Weight
Net: 13 kg (28.5 lbs.).
Shipping: 16 kg (35.2 lbs.).
Dimensions (excluding handle)
Height: 216 mm (8.5 in).
Width: 437 mm (17.19 in).
Depth: 440 mm (17.34 in).
Safety
Meets IEC 61010-1 +A2, CSA certified to C22.2 No.1010.1, self-certified to UL 3111.
* Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period, and ±5°C from annual calibration temperature.
1 Full scale is defined as 8 vertical divisions. Vertical divisions are defined by the major scale settings above non-major scale settings. The major scale settings are 10 mV, 20 mV,
50 mV, 100 mV, 200 mV, 500 mV, 1 V.
2 Vertical resolution for 8 bits = 0.4% of full scale, for 12 bits = 0.024% of full scale.
3 Within one year of previous calibration.
4 FFT amplitude readings are affected by input amplifier roll-off. 54855A: (–3 dB at 6 GHz, with amplitude decreasing as frequency increases above 6 GHz).
54854A: (–3 dB at 4 GHz, with amplitude decreasing as frequency increases above 4 GHz).
5 The noise floor varies with memory depth and averaging.
6 Signal peak-to-peak amplitude
≥
5 divisions, vertical scale
≥
10 mV/div, signal rise time
≤
225 ps (54854) 150 ps (54855), sample rate = 20 GSa/s, sin(x)/x interpolation enabled,
measurement threshold = fixed voltage at 50 % level.
7 Between two edges on a single channel. Rms value refers to the standard deviation of 256 consecutive measurements performed using an individual instrument.
8 Internal trigger. Trigger level contained within full scale display range of trigger channel.