Continued) – Atec Agilent-54855A User Manual
Page 11
11
www.agilent.com/find/infiniimax
Infiniium 54850 Series Performance Characteristics
(continued)
Acquisition
Real time sample rate per channel
20 GSa/s
Memory depth per channel
Standard
262,144 at all sample rates
Option 001
1,025,000 at all sample rates
32,800,000
≤
2 GSa/s sample rate
Sampling modes
Real time
Successive single-shot acquisitions
Real time with averaging
Selectable from 2 to 4096
Real time with peak detect
2 GSa/s peak detect, for less than 2 GSa/s sample rates (option 001 only)
Filters
Sin(x)/x Interpolation
On/off selectable FIR digital filter. Digital signal processing adds points between
acquired data points to enhance measurement accuracy and waveform display quality.
Trigger
Sensitivity
1
Internal Low
1
54855A: 0.5 div p-p 0 to 2 GHz, 1.0 div p-p 2 to 4 GHz, < 2.5 div @ 5 GHz
54854A: 0.5 div p-p 0 to 2 GHz, 1.0 div p-p 2 to 4 GHz
54853A: 0.5 div p-p 0 to 2 GHz, 1.0 div p-p 2 to 2.5 GHz
Internal High
1
54855A: 0.2 div p-p 0 to 6 GHz
54854A: 0.2 div p-p 0 to 4 GHz
54853A: 0.2 div p-p 0 to 2.5 GHz
Auxiliary
DC to 500 MHz: 500 mV p-p
Level range
Internal
± 8 div from center screen or ± 4 Volts, whichever is smallest
Auxiliary
± 5 V
Sweep modes
Auto, triggered, single
Trigger jitter
6,8
54855A: 1.0 ps rms
54854A: 1.3 ps rms
54853A: 1.7 ps rms
Trigger holdoff range
80 ns to 320 ms
Trigger actions
Specify an action to occur and the frequency of the action when a trigger
condition occurs. Actions include e-mail on trigger and QuickMeas+.