beautypg.com

Specifications – Atec Agilent-4284A_4285A User Manual

Page 20

background image

20

q

accuracy is given as:

±[(180/p) x (A

e

/100)] (absolute degrees)

Note: 1.

A

e

= (A

n

+ B)

Additional error due to temperature:

Multiply the measurement accuracy by the following temperature factors.

Example L and Q accuracy calculation

Measurement conditions:

Frequency:

500 kHz

L value measured:

2 mH

Test signal level:

1 Vrms

Integration time:

Long

Cable length:

0 meters

Q value measured (Qx): 200

Calculation:

Step 1:

Use Figure 2 to determine A

n

and Z

m

.

a. Find the frequency along the X-axis.
b. Find the inductance value along a diagonal.
c. Note the intersection of steps a and b.

In this case A

n

= A

5

. Refer to the equations in Table 3.

d. Note that in step c Z

m

is 6.3 k

Ω.

Step 2:

Use Tables 3 and 4 to determine A

n

and B.

a. A

n

is equation A

5

for long integration times:

0.18% + [(|Z

m

|/5 k) x 0.02%]

b. A

5

yields a value of 0.21%

c. Table 4 indicates that B has a value of 0.

(@ cable length = 0 m)

d. L accuracy is ±(A

n

+ B) = 0.21%

e. Determine D accuracy (D

e

): (A

n

+ B)/100 = 0.0021

f. Q accuracy: (

∆Q)

±[(Q

x

2

x D

e

)/(1 ] (Q

x

x D

e

)]

g.

∆Q yields a value of –57 to 133, Actual Q: 143 to 333
N

1

, N

2

and N

3

are in Table 3.

Table 4. Cable length correction

Test cable length

B (%)

0 meter

0

1 meter (16048A)

f

m

/15

2 meter (16048D)

f

m

/15

(fm : test frequency in MHz)

Specifications continued on page 21

Specifications

Continued

from page 19