Specifications, Measurement range, Display – Atec Agilent-4284A_4285A User Manual

Page 11: Correction function, List sweep function

background image

11

Agilent 4285A: standard

Range

Accuracy

Normal

V

5 m Vrms to 2 Vrms

±(8% + 0.4 fm% + 1 mVrms)

I

200 µArms to 20 mArms

±(8% + 1 fm% 40 µArms)

Constant

V

10 m Vrms to 1 Vrms

±(6% + 0.2 fm%+ 1 mVrms)

I

100 µArms to 20 mArms

±(6% + 0.2 fm%+ 40 µArms)

fm: test frequency in MHz

DC bias:

Standard:

0 V, 1.5 V and 2 V (Agilent 4284A only)

With
Option 4284A/4285A-001: 0 V to ±40 V Rear panel DC bias monitor, BNC connector

Range

Resolution

Accuracy

±(0.000 to 4.000) V

1 mV

±(0.1% to 1 mV)

1

±(4.002 to 8.000) V

2 mV

±(0.1% to 2 mV)

1

±(8.005 to 20.000) V

5 mV

±(0.1% to 5 mV)

1

±(20.01 to 40.00) V

10 mV

±(0.1% to 10 mV)

1

Measurement range

Parameter Range
|Z|, R, X:

0.01 m

Ω to 99.9999 MΩ

|Y|, G, B:

0.01 nS to 99.9999 S

C:

4284A: 0.01 fF to 9.99999 F
4285A: 0.01 fF to 999.999 µF

L:

4284A: 0.01 nH to 99.9999 kH
4285A: 0.001 nH to 99.9999 H

D:

0.000001 to 9.99999

Q:

0.01 to 99999.9

q

:

-180.000° to 180.000°

∆%

-999.999% to 999.999%

Display

LCD Dot-matrix type display. Capable of displaying: measured values, control
settings, comparator limits and decisions, list sweep tables, self test messages
and annunciations.

Correction function

Open/short:

Eliminates measurement errors due to stray parasitic
impedance in the test fixture.

Load:

Improves measurement accuracy by using a calibrated
device as a reference.

List sweep function

A maximum of ten frequencies or test levels can be programmed. Single or
sequential testing can be performed. When Option 4284A/4285A-001 is installed,
dc bias sweep can also be performed.

Specifications continued on page 12

Specifications

Continued

from page 10

1. Refer to measurement accuracy.