Specifications, Measurement range, Display – Atec Agilent-4284A_4285A User Manual
Page 11: Correction function, List sweep function
![background image](/manuals/318038/11/background.png)
11
Agilent 4285A: standard
Range
Accuracy
Normal
V
5 m Vrms to 2 Vrms
±(8% + 0.4 fm% + 1 mVrms)
I
200 µArms to 20 mArms
±(8% + 1 fm% 40 µArms)
Constant
V
10 m Vrms to 1 Vrms
±(6% + 0.2 fm%+ 1 mVrms)
I
100 µArms to 20 mArms
±(6% + 0.2 fm%+ 40 µArms)
fm: test frequency in MHz
DC bias:
Standard:
0 V, 1.5 V and 2 V (Agilent 4284A only)
With
Option 4284A/4285A-001: 0 V to ±40 V Rear panel DC bias monitor, BNC connector
Range
Resolution
Accuracy
±(0.000 to 4.000) V
1 mV
±(0.1% to 1 mV)
1
±(4.002 to 8.000) V
2 mV
±(0.1% to 2 mV)
1
±(8.005 to 20.000) V
5 mV
±(0.1% to 5 mV)
1
±(20.01 to 40.00) V
10 mV
±(0.1% to 10 mV)
1
Measurement range
Parameter Range
|Z|, R, X:
0.01 m
Ω to 99.9999 MΩ
|Y|, G, B:
0.01 nS to 99.9999 S
C:
4284A: 0.01 fF to 9.99999 F
4285A: 0.01 fF to 999.999 µF
L:
4284A: 0.01 nH to 99.9999 kH
4285A: 0.001 nH to 99.9999 H
D:
0.000001 to 9.99999
Q:
0.01 to 99999.9
q
:
-180.000° to 180.000°
∆%
-999.999% to 999.999%
Display
LCD Dot-matrix type display. Capable of displaying: measured values, control
settings, comparator limits and decisions, list sweep tables, self test messages
and annunciations.
Correction function
Open/short:
Eliminates measurement errors due to stray parasitic
impedance in the test fixture.
Load:
Improves measurement accuracy by using a calibrated
device as a reference.
List sweep function
A maximum of ten frequencies or test levels can be programmed. Single or
sequential testing can be performed. When Option 4284A/4285A-001 is installed,
dc bias sweep can also be performed.
Specifications continued on page 12
Specifications
Continued
from page 10
1. Refer to measurement accuracy.