Basic measurement characteristics – Atec Agilent-4287A User Manual
Page 7
Frequenc y > 1.8 GHz,
≤ 3 GHz
± 5 [%]
Oscillator level < –33 dBm
Frequenc y
≥ 1 MHz,
≤ 100 MHz
± 1 [%]
Frequenc y > 100 MHz,
≤ 500 MHz
± 1.2 [%]
Frequenc y > 500 MHz,
≤ 1 GHz
± 1.2 [%]
Frequenc y > 1 GHz,
≤ 1.8 GHz
± 2.5 [%]
Frequenc y > 1.8 GHz,
≤ 3 GHz
± 5 [%]
E
b
=
[%]
(| Z
x
|: Measurement value of |Z|)
E
c
=
[%] ( F: Frequency [MHz])
Z
s
=
(Within
± 5 °C from the calibration temperature.
Measurement accuracy applies when the calibration
is performed at 23
°C ± 5 °C. When the calibration is
performed beyond 23
°C ± 5 °C, the measurement
accuracy decreases to half that described.)
Oscillator level
≥ –33 dBm,
Averaging factor
≥ 8
± (20 + 0.5 × F ) [mΩ] (F: Frequency [MHz])
Oscillator level
≥ –33 dBm,
Averaging factor < 7
± (50 + 0.5 × F ) [mΩ] (F: Frequency [MHz])
± (100 + 0.5 × F ) [m Ω] (F: Frequency [MHz])
Y
o
=
(Within
± 5 °C from the calibration temperature.
Measurement accuracy applies when the calibration
is performed at 23
°C ± 5 °C. When the calibration is
performed beyond 23
°C ± 5 °C, the measurement
accuracy decreases to half that described.)
± (30 + 0.15 × F ) [µS] ( F: Frequency [MHz])
± (50 + 0.15 × F ) [µS] ( F: Frequency [MHz])
± (100 + 0.15 × F ) [µS] ( F: Frequency [MHz])
Z
s
Z
x
--------
Y
o
Z
x
•
+
100
×
±
0.06
0.08
F
×
1000
-------------------
+
±
(
)
(
)
Oscillator level
≥ –33 dBm,
Oscillator level
≥ –33 dBm,
Averaging factor
≥ 8
Oscillator level
≥ –33 dBm,
Averaging factor < 7
Oscillator level
≥ –33 dBm,
7
Basic Measurement Characteristics