Measurement support functions – Atec Agilent-4287A User Manual

Page 13

background image

DC resistance (Rdc) measurement

Trigger function

Averaging function

13

Measurement range

0.1

Ω to 100 Ω

Measurement resolution

1 m

Test signal level

1 mA (maximum)

Error correction

OPEN/SHOR T/LO AD Calibration, OPEN/SHORT
Compensation. (Changing the frequency or oscillator
level settings after the calibration or compensation
makes the calibration and compensation data invalid.)

Measurement uncertainty

[%]

R

dut

: DC resistance measurement v alue [

Ω]

(Within

± 5 °C from the calibration temperature.

Measurement accuracy applies when the calibration
is performed at 23

°C ± 5 °C. When the calibration is

performed beyond 23

°C ± 5 °C, the measurement

accuracy decreases to half that described.)

Trigger mode

Internal, External (external trigger input connector or
handler interface), Bus (GPIB or LAN), Manual (front key)

Setting range

1 to 100 (integer)

1

0.05
R

dut

---------

R

dut

10000

--------------

+

100

×

+

±

( )

Measurement Support Functions