Measurement support functions – Atec Agilent-4287A User Manual
Page 12
12
OPEN/SHORT/LOAD
Calibration
Connect OPEN, SHORT, and LOAD standards to the
desired reference plane and measure each kind of
calibration data. The reference plane is called calibration
reference plane.
Low-Loss Capacitor
Calibration
Connect the dedicated standard (Low-Loss Capacitor) to the
calibration reference plane and measure the calibration data.
Port Extension
Compensation
(Fixture Selection)
When a device is connected to the terminal that is extended
from the calibration reference plane, set the electrical length
between the calibration plane and the device contact. Select
a model number of the registered test fixtures in the 4287A's
softkey menu or enter the electrical length for user's test
fixture.
OPEN/SHORT
Compensation
When a device is connected to the terminal that is extended
from the calibration reference plane, make OPEN and
SHORT states at the device contact and measure each kind
of compensation date.
Data Measurement
Points
Same as measurement points which is set in the
measurement point setup display. (Changing the frequency
or oscillator level settings after the calibration or
compensation makes the calibration and compensation data
invalid.)
Measurement Support Functions
Error correction function
Available calibration and compensation
Calibration/compensation data measurement point