Basic measurement characteristics – Atec Agilent-4287A User Manual

Page 4

background image

Range

Cable length: 1m

Power (when 50

Ω LOAD is

connected to the test port)

–40 dBm to 1 dBm (Frequency

≤ 1 GHz)

–40 dBm to 0 dBm (Frequency > 1 GHz )

*1

Current (when SHORT is
connected to the test port)

0.0894 mA

rms

to 10 mA

rms

(Frequency

≥ 1 GHz)

0.0894 mA

rms

to 8.94 mA

rms

(Frequency > 1 GHz )

Voltage (when OPEN is
connected to the test port)

4.47 mV

rms

to 502 mV

rms

4.47 mV

rms

to 447 mV

rms

Cable length: 2m

(when Option 4287A-020 is used)

Power

Subtract the following attenuation from the power
(setting value) at 1 m cable length:

Attenuation [dB] = 0.37

×

(F: Frequency [GHz])

Resolution

0.1 dB

*2

Uncertainty

Cable length: 1 m

Power (when 50

Ω LOAD is

connected to the test port)

Frequency

≤ 1 GHz

± 2 dB (23 ± 5 ºC)

± 4 dB (5 ºC to 40 ºC)

Frequency > 1 GHz

± 3 dB (23 ± 5 ºC)

± 5 dB (5 ºC to 40 ºC)

Cable length: 2 m

(when Option 4287A-020 is used)

Power

Add 1 dB to the uncertainty at 1 m cable length.

Output impedance

50

Ω (nominal)

F

*1

(Frequency

≥ 1 GHz)

(Frequency > 1 GHz )

*1

4

Basic Measurement Characteristics

Oscillator level

Output Impedance

*1. It is possible to set more than 0 dBm (447 mV, 8.94 mA) oscillator level at

frequency > 1 GHz. However, the characteristics at this setting are not guaranteed.

*2. When the unit is set at mV or mA, the entered value is rounded to 0.1 dBm resolution.