Dc bias settling time – Atec Agilent-E4980A-AL User Manual

Page 36

background image

36

Table 66. Yo_DCI1 value

DC bias current range

Measurement time mode

SHORT

MED,

LONG

20 µA

0 S

0 S

200 µA

0.25 nS

0.05 nS

2 mA

2.5 nS

0.5 nS

20 mA

25 nS

5 nS

100 mA

250 nS

50 nS

Table 67. Yo_DCI2 value

DC bias

Measurement time mode

current range

≤ 100 Ω

300 Ω, 1 k Ω

3 k Ω, 10 k Ω

30 k Ω, 100 k Ω

20 µA

0 S

0 S

0 S

0 S

200 µA

0 S

0 S

0 S

0 S

2 mA

0 S

0 S

0 S

3 nS

20 mA

0 S

0 S

30 nS

30 nS

100 mA

0 S

300 nS

300 nS

300 nS

DC bias settling time

When DC bias is set to ON, add the following value to the settling time:

Table 68. DC bias settling time

Bias

Settling

time

1

Standard

Capacitance of DUT Ч 100 Ч log

e

(2/1.8 m) + 3 m

2

Option 001

Capacitance of DUT Ч 100 Ч log

e

(40/1.8 m) + 3 m

Figure 12. DC bias settling time

10 msec

100 msec

1 sec

10 sec

100 sec

1.

2.

1 µF

10 µF

100 µF

1 mF

10 mF

100 mF

DUT capacitance

Settling time