beautypg.com

Power and dc bias enhancement specification, Test signal – Atec Agilent-E4980A-AL User Manual

Page 20

background image

20

Power and DC Bias enhancement specification

Increases test signal voltage and adds the variable DC bias voltage function.
This function is available when the option 001 is installed.

Measurement parameters

The following parameters can be used.

• Lp-Rdc
• Ls-Rdc
• Vdc-Idc

where

Rdc Direct-current resistance (DCR)
Vdc Direct-current

voltage

Idc Direct-current

electricity

Test signal

Signal level

Table 31. Test signal voltage

Range

0 Vrms to 20 Vrms (test frequency ≤ 1 MHz)

0 Vrms to 15 Vrms (test frequency > 1 MHz)

Resolution

100 µVrms (0 Vrms - 0.2 Vrms)

200 µVrms (0.2 Vrms - 0.5 Vrms)

500 µVrms (0.5 Vrms - 1 Vrms)

1 mVrms (1 Vrms - 2 Vrms)

2 mVrms (2 Vrms - 5 Vrms)

5 mVrms (5 Vrms - 10 Vrms)

10 mVrms (10 Vrms - 20 Vrms)

Setup accuracy

normal

±(10% + 1 mVrms) (test signal voltage ≤ 2 Vrms)

(test frequency ≤ 1 MHz : spec., test frequency > 1 MHz : typ.)

±(10% + 10 mVrms) (Test frequency ≤ 300 kHz,

test signal voltage > 2 Vrms) (spec.)

±(15% + 20 mVrms) (test frequency > 300 kHz,

test signal voltage > 2 Vrms) (test frequency ≤ 1 MHz : spec.,

test frequency > 1 MHz : typ.)

Constant

1

±(6% + 1 mVrms) (test signal voltage ≤ 2 Vrms)

(test frequency ≤ 1 MHz : spec. , test frequency > 1 MHz : typ.)

±(6% + 10 mVrms) (test frequency ≤ 300 kHz,

test signal voltage > 2 Vrms) (spec.)

±(12% + 20 mVrms) (test frequency > 300 kHz,

test signal voltage > 2 Vrms) (test frequency ≤ 1 MHz : spec.,

test frequency > 1 MHz : typ.)

1. When auto level control function is on.