DCR measurement specification
DC resistance (Rdc) measurement function is available when either option 001, 030, 050,
100, or 200 is installed.
DC resistance (Rdc) accuracy
Absolute measurement accuracy Aa
Absolute measurement accuracy Aa is given as
Equation 15.
Aa = Ae + Acal
Aa
Absolute accuracy (% of reading value)
Ae
Relative accuracy (% of reading value)
Acal Calibration accuracy
Relative measurement accuracy Ae
Relative measurement accuracy Ae is given as
Equation 16.
Ae = [Ab + (Rs /|Rm|+ Go × |Rm|) Ч 100 ] Ч Kt
Rm Measurement
value
Ab Basic
accuracy
Rs
Short offset [Ω]
Go Open
offset
[S]
Kt
Temperature coefficient
Calibration accuracy Acal
Calibration accuracy Acal is 0.03%.
Basic accuracy Ab
Table 43. Basic accuracy Ab is given below.
Measurement
Test signal voltage
time mode
≤ 2 Vrms
> 2 Vrms
SHORT
1.00%
2.00%
MED
0.30%
0.60%
Open offset Go
Table 44. Open offset Go is given below.
Measurement
Test signal voltage
time mode
≤ 2 Vrms
> 2 Vrms
SHORT
50 nS
500 nS
MED
10 nS
100 nS
Short offset Rs
Table 45. Short offset Rs is given below.
Measurement
Test signal voltage
time mode
≤ 2 Vrms
> 2 Vrms
SHORT
25 mΩ
250 mΩ
MED
5 mΩ
50 mΩ