DCR measurement specification

DC resistance (Rdc) measurement function is available when either option 001, 030, 050,
100, or 200 is installed.

DC resistance (Rdc) accuracy

Absolute measurement accuracy Aa

Absolute measurement accuracy Aa is given as

Equation 15.

Aa = Ae + Acal

Aa

Absolute accuracy (% of reading value)

Ae

Relative accuracy (% of reading value)

Acal Calibration accuracy

Relative measurement accuracy Ae

Relative measurement accuracy Ae is given as

Equation 16.

Ae = [Ab + (Rs /|Rm|+ Go × |Rm|) Ч 100 ] Ч Kt

Rm Measurement

value

Ab Basic

accuracy

Rs

Short offset [Ω]

Go Open

offset

[S]

Kt

Temperature coefficient

Calibration accuracy Acal

Calibration accuracy Acal is 0.03%.

Basic accuracy Ab

Table 43. Basic accuracy Ab is given below.

Measurement

Test signal voltage

time mode

≤ 2 Vrms

> 2 Vrms

SHORT

1.00%

2.00%

MED

0.30%

0.60%

Open offset Go

Table 44. Open offset Go is given below.

Measurement

Test signal voltage

time mode

≤ 2 Vrms

> 2 Vrms

SHORT

50 nS

500 nS

MED

10 nS

100 nS

Short offset Rs

Table 45. Short offset Rs is given below.

Measurement

Test signal voltage

time mode

≤ 2 Vrms

> 2 Vrms

SHORT

25 mΩ

250 mΩ

MED

5 mΩ

50 mΩ