beautypg.com

Atec Agilent-4395A User Manual

Page 9

background image

Linear scale

2

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

<±3%

1

Fidelity shows an extent of nonlinearity referenced to the reference input level.

2

RBW = 10 Hz, –20 dBm

reference value

+30 dBm, reference input level=full scale input level –10 dB, 23 ± 5°C

Note: Refer to Input attenuator part for the definition of full scale input level.

Displayed average noise level

at reference value

–40 dBm, input attenuator: auto or 0 dB

at frequency

1 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –120 dBm/Hz

at

100 kHz

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –133 dBm/Hz

at

10 MHz

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (–145 + frequency/100 MHz) dBm/Hz

1

1

at start frequency

10 MHz

Figure 1-4. Typical Displayed Average Noise Level

9