beautypg.com

Atec Agilent-4395A User Manual

Page 5

background image

Residual responses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . <–80 dB full scale (SPC)
Trace noise (A/R, B/R, A/B)

at 50 MHz, both inputs: full scale input level –10 dB, IFBW = 300 Hz . . . . . . . . . . . . <0.005 dB rms (SPC)

Stability (A/R, B/R, A/B) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < ±0.01 dB/°C (SPC)

Phase Characteristics

Measurements format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Standard format, Expanded phase format
Frequency response (deviation from linear phase) (A/R, B/R) (SPC for A/B)

at –10 dBm input, input attenuator: 10 dB, IFBW

3 kHz, 23 ± 5°C . . . . . . . . . . . . . . . . . . . . . . . . . < ±12°

Dynamic accuracy (A/R, B/R) (SPC for A/B)

Input Level

(relative to full scale input level –10 dB)

Dynamic Accuracy

1

frequency

100 Hz

0 dB

input level > –10 dB

±3°

–10 dB

input level

–60 dB

±0.3°

–60 dB > input level

–80 dB

±1.8°

–80 dB > input level

–100 dB

±18°

1. R input level (B input level for A/B) = full scale input level –10 dB, IFBW = 10 Hz, 23 ± 5°C

Figure 1-2. Phase Dynamic Accuracy

Trace noise (A/R, B/R, A/B)

at 50 MHz, both inputs: full scale input level –10 dB, IFBW=300 Hz . . . . . . . . . . . . . . . . <0.04° rms (SPC)

Stability (A/R, B/R, A/B) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

<

±0.1 °/°C (SPC)