Atec Agilent-4395A User Manual
Page 25
S-parameter test set interface
Connector
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-SUB (25-pin)
Caution
Do not connect a printer to this connector. If you connect a printer with the S-parameter
test set interface connector (TEST SET-I/O INTERCONNECT), it may cause damage to
the printer.
Figure 1-11. S-Parameter Test Set Interface Pin Assignments
External monitor output
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
D-SUB (15-pin HD)
Display resolution
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 640 x 480 VGA
Operation Conditions
Temperature
Disk drive non-operating condition
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0°C to 40°C
Disk drive operating condition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10°C to 40°C
Humidity
at wet bulb temperature
≤
29°C, without condensation
Disk drive non-operating condition
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15% to 95% RH
Disk drive operating condition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15% to 80% RH
Altitude
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 2,000 m
Warm up time
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 minutes
25