Specifications describe the instrument's warranted performance over the temperature range of 0°C to 40°C
(except as noted). Supplemental characteristics are intended to provide information that is useful in applying
the instrument by giving non-warranted performance parameters. These are denoted as SPC (supplemental
performance characteristics)

, typical, or nominal. Warm up time must be greater than or equal to 30 minutes

after power on for all specifications.

Network Measurement

Source Characteristics

Frequency Characteristics

Range

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Hz to 500 MHz

Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 mHz
Frequency reference

Accuracy

at 23 ± 5°C, referenced to 23°C

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . <±5.5 ppm

Aging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . <±2.5 ppm/year (SPC)
Initial achievable accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . <±1.0 ppm (SPC)
Temperature stability

at 23 ± 5°C, referenced to 23°C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . <±2ppm (SPC)

Precision frequency reference (option 1D5)
Accuracy

at 0°C to 40°C, referenced to 23°C

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . <±0.13 ppm

Aging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . <±0.1 ppm/year (SPC)

Initial achievable accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . <±0.02 ppm (SPC)
Temperature stability

at 0°C to 40°C, referenced to 23°C

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . <±0.01 ppm (SPC)

Output Characteristics

Power range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –50 dBm to +15 dBm
Level accuracy

at 0 dBm output, 50 MHz, 23 ± 5°C, . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±1.0 dB

Level linearity

Output Power

Linearity

1

–40 dBm

±1.0 dB

<

–40 dBm

±1.5 dB

1 at relative to 0 dBm output, 50 MHz, 23 ± 5°C

HP 4395A
Network/Spectrum/Impedance Analyzer

Technical Specifications