On-screen dynamic range
Figure 1-5. Typical On-screen Dynamic Range (Center: 100 MHz)
Spurious responses
Second harmonic distortion
at single tone input with full scale input level –10 dB, input signal frequency
≥
100 kHz
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . <–70 dBc, <–75 dBc (SPC)
Third order inter-modulation distortion
at two tones input with full scale input level –16 dB, separation
≥
100 kHz
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . <–75 dBc, <–80 dBc (SPC)
Spurious
at single tone input with full scale input level –10 dB, input signal frequency
≤
500 MHz
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . <–75 dBc
except for the following frequency ranges:
5.6 MHz ±1 MHz, 30.6 MHz ±1 MHz, 415.3 MHz ±1 MHz
Residual response
at reference value setting
≤
–40 dBm, input attenuator: auto or 0 dB
. . . . . . . . . . . . . . . . . . . . <–110 dBm
10