Atec Agilent-E7495A User Manual
Page 3
![background image](/manuals/317961/3/background.png)
Chapter 21
627
Safety and Regulatory Information and Specifications
E7495A/B Base Station Test Set Specifications
Safety and Regula
tory In
fo
rmatio
n
a
n
d
Sp
ecification
s
Amplitude accuracy:
±1 dB (100 MHz – 2500 MHz at 25 °C)
Resolution bandwidth:
10 Hz to 1 MHz, settable to 1 Hz precision
Span:
1 kHz to 2.6995 GHz
Trace update:
Span:
2.6995 GHz (autocouple) = 5.1 seconds
60 MHz (autocouple) = 400 ms
1 MHz (100 Hz RBW) = 1.2 seconds
Simultaneous dynamic range:
>
90 dB (CW signals at 300 kHz separation, span 500 kHz,
30 Hz RBW
SSB phase noise:
≤
–85 dBc (30 kHz offset)
Spurious responses:
Range control set to auto, high
sensitivity mode internally generated.
50 ohm load on input:
Crossing spurs:
<
115 dBm
≤
50 dBc
Displayed average noise level:
–150 dBm (30 Hz RBW, 375 MHz to 1.5 GHz)
Port 2 VSWR:
<
2:1
Antenna/Cable Tester
For each of the following measurements, a short self-calibration procedure must be run prior to making the
measurement.
Frequency range:
375 to 2500 MHz
Frequency resolution:
< 500 Hz
Immunity to interfering signals:
+20 dBm (with interference rejection turned on)
Measurement speed:
Full span:
60 MHz span:
<
17 ms/point
<
7 ms/point
Table 21-1
E7495A/B Base Station Test Set General Specifications