background image

626

Chapter 21

Safety and Regulatory Information and Specifications
E7495A/B Base Station Test Set Specifications

Safety and Regulatory Information

a

n

d

Sp

ecification

s

Maximum input power without damaging
instrument:

100 W (without external attenuator)

Frequency and time reference:

Can use internal timebase or external
signal:

GPS (external antenna supplied)
Even second; pulse

1.000 MHz:
2.048 MHz:
4.950 MHz:
10.000 MHz:
13.000 MHz:
15.000 MHz:
19.6608 MHz:

0 dBm (E7495B only)

0 dBm (E7495B only)

0 dBm (E7495B only)

0 dBm

0 dBm (E7495B only)

0 dBm (E7495B only)

0 dBm

Display:

Scale:
Number of points:
Number of divisions:

1 to 20 dB / div. Settable in 1 dB increments.
256
10

40 dB Attenuator:

Frequency range:
Attenuation accuracy:
Max power:

10 to 2500 MHz
±0.5 dB
50 dBm (100 W)

Spectrum Analyzer / Tools

Input frequency range:

10 mHz to 2700 MHz (usable to 500 kHz)

Reference level range:

–150 to +100 dBm

Dynamic range:

+50 dBm to –150 dBm
(with supplied external 40 dB attenuator) (30 Hz RBW)

Input attenuation:

0 to 30 dB automatically selected.
10 dB controllable manually.

Table 21-1

E7495A/B Base Station Test Set General Specifications