Atec Agilent-E7495A User Manual
Page 2

626
Chapter 21
Safety and Regulatory Information and Specifications
E7495A/B Base Station Test Set Specifications
Safety and Regulatory Information
a
n
d
Sp
ecification
s
Maximum input power without damaging
instrument:
100 W (without external attenuator)
Frequency and time reference:
Can use internal timebase or external
signal:
GPS (external antenna supplied)
Even second; pulse
1.000 MHz:
2.048 MHz:
4.950 MHz:
10.000 MHz:
13.000 MHz:
15.000 MHz:
19.6608 MHz:
≥
0 dBm (E7495B only)
≥
0 dBm (E7495B only)
≥
0 dBm (E7495B only)
≥
0 dBm
≥
0 dBm (E7495B only)
≥
0 dBm (E7495B only)
≥
0 dBm
Display:
Scale:
Number of points:
Number of divisions:
1 to 20 dB / div. Settable in 1 dB increments.
256
10
40 dB Attenuator:
Frequency range:
Attenuation accuracy:
Max power:
10 to 2500 MHz
±0.5 dB
50 dBm (100 W)
Spectrum Analyzer / Tools
Input frequency range:
10 mHz to 2700 MHz (usable to 500 kHz)
Reference level range:
–150 to +100 dBm
Dynamic range:
+50 dBm to –150 dBm
(with supplied external 40 dB attenuator) (30 Hz RBW)
Input attenuation:
0 to 30 dB automatically selected.
10 dB controllable manually.
Table 21-1
E7495A/B Base Station Test Set General Specifications