beautypg.com

Atec Agilent-81624B User Manual

Page 9

background image

9

1. Insertion Loss is in the range of 7 dB
2. Averaging time 1 s

• calibration prior to measurement
• constant temperature
• broadband source: Agilent 83438A
• FP Sources: Agilent 81650A, 81651A, 81654A with active Coherence Control
• Reference Cable 81610CC used for total uncertainty
• Length of measurement patch cord ≤ 2 m, angled connector in optimal optical conditions

3. Warm-up time 60 min, if previously not stored at the same temperature

Agilent 81610CC Reference cable

Return loss

As printed on cable

Return loss uncertainty

±0.2 dB

1

Wavelengths

1310 nm and 1550 nm ±15 nm

1. Clean reference reflector in perfect optical condition (Do not use with contact-type connectors)

Return loss module specifications

All modules require angled contact (8°) at input and output connectors

Agilent 81610A

Source

External input only

1

Sensor element

InGaAs

Fiber type

Standard single-mode 9/125 μm

External input

Max input power:

10 dBm

Min input power:

0 dBm

Damage input power:

16 dBm

Wavelength range for external input

1250 nm to 1640 nm

Dynamic range

70 dB

Relative uncertainty of

2

Return Loss (RL)

RL ≤ 55 dB

RL ≤ 60 dB

RL ≤ 65 dB

RL ≤ 70 dB

Total uncertainty

with broadband source

< ±0.25 dB

< ±0.3 dB

< ±0.65 dB

< ±1.7 dB

Add ±0.2 dB

with Agilent FP sources

Typ. < ±0.5 dB

Typ. < ±1.0 dB

Typ. < ±2.0 dB

Add typ. ±0.2 dB

Maximum safe input power

+16 dBm

Dimensions (H x W x D)

75 mm x 32 mm x 335 mm (2.8” x 1.3” x 13.2”)

Weight

0.6 kg

Recommended recalibration period

2 years

Operating temperature

10 °C to 40 °C

Humidity

Non-condensing

Warm-up time

3

20 minutes

Reference Cable Specification

To connect to Return Loss Modules the cable requires connector Interface 81000SI DIN47256/4108