beautypg.com

Atec Agilent-81624B User Manual

Page 5

background image

5

Fast power sensor module specifications

Agilent 81636B

Sensor element

InGaAs

Wavelength range

1250 nm to 1640 nm

Power range

–80 dBm to +10 dBm

Applicable fiber type

Standard SM and MM up to 62.5 μm core size, NA ≤ 0.24

Uncertainty (accuracy) at reference conditions

1

±3% (1260 nm to 1630 nm)

Total uncertainty

2, 9

±5% ±20 pW

8

(1260 nm to 1630 nm)

Relative uncertainty:

- due to polarization

3

- spectral ripple (due to interference)

4

Typ. ±0.015 dB

Typ. ±0.015 dB

Linearity (power)

5, 9

- at 23 °C ±5 °C

- at operating temperature range

CW –60 to +10 dBm, (1260 nm to 1630 nm)

< ±0.02 dB

< ±0.06 dB

Return loss

7

> 40 dB

Noise (peak to peak)

6

< 20 pW (1260 nm - 1630 nm)

Averaging time (minimal)

25 μs

Dynamic Range at manual range mode

5, 10

- at +10 dBm-range

- at ±0 dBm-range

- at –10 dBm-range

- at –20 dBm-range

Typ. > 55 dB

Typ. > 55 dB

Typ. > 52 dB

Typ. > 45 dB

Noise at manual range mode (peak to peak):

10

- at +10 dBm-range

- at ±0 dBm-range

- at –10 dBm-range

- at –20 dBm-range

CW –60 to +10 dBm, 1260 nm to 1630 nm

< 50 nW

< 5 nW

< 1 nW

< 500 pW

Analog output

Included

Maximum safe input power

+16 dBm

Dimensions (H x W x D)

75 mm x 32 mm x 335 mm (2.8” x 1.3” x 13.2”)

Weight

0.5 kg

Recommended recalibration period

2 years

Operating temperature

+10 °C to +40 °C

Humidity

Non-condensing

Warm-up time

20 min

1. Reference Conditions:

• Power level 10 μW (-20 dBm),

continuous wave (CW)

• Fiber 50 μm graded-index, NA = 0.2
• Ambient temperature 23 °C ±5 °C
• On day of calibration (add ±0.3% for aging

over one year, add ±0.6% over two years)

• Spectral width of source < 10 nm (FWHM)
• Wavelength setting at power sensor must

correspond to source wavelength ±0.4 nm

2. Operating Conditions:

• Fiber ≤ 50 μm, NA ≤ 0.2
• For LC connector use 81000LI adapter, not

81002LI; for MU connector use 81000MI,

not 81002MI

• Averaging time 1 s
• Within one year after calibration,

add 0.3% for second year

• Operating temperature range as specified

humidity: none condensing

3. All states of polarization at constant wave-

length (1550 nm ±30 nm) and constant

power, straight connector, T = 23 °C ±5 °C.

For angled connector (8°) add ±0.01 dB typ.
4. Conditions:

• Wavelength 1550 nm ±30 nm, fixed

state of polarization, constant power

• Temperature 23 °C ±5 °C
• Linewidth of source ≥ 100 MHz
• Angled connector 8°

5. Does not include noise
6. Averaging time 1 s, T = 23 °C ±5 °C,

ΔT ±1 °C, observation time 300 s
7. Conditions:

• Wavelengths

1310 nm ±30 nm and 1550 nm ±30 nm

• Standard single mode fiber
• Angled connector min 8°. T = 23 °C ±5 °C

8. For wavelengths > 1600 nm add ±0.06%/nm
9. For input power > 2 mW add ±0.02 dB
10. Conditions:

• Averaging time 25 μs, T = 23 °C ±5 °C
• Observation time 2.5 s