81624B high performance and 81626B high power optical head specifications

All optical heads have to be operated with the single (Agilent 81618A) or dual (Agilent 81619A) Interface Modules

Agilent 81624B

Agilent 81624B

Calibration option

C01/C02

Agilent 81626B

Agilent 81626B

Calibration option

C01/C02

Sensor element

InGaAs, Ø5 mm

InGaAs, Ø5 mm

Wavelength range

800 nm to 1700 nm

850 nm to 1650 nm

Power range

–90 dBm to +10 dBm

–70 to +27 dBm (1250 nm to 1650 nm)

–70 to +23 dBm (850 nm to 1650 nm)

Applicable fiber type

Open beam

Standard SM and MM max 100 µm core size, NA ≤ 0.3

Parallel beam max Ø4 mm

Standard SM and MM max 100 µm core size, NA ≤ 0.3

Parallel beam max Ø4 mm

Uncertainty (accuracy) at

reference conditions

1

±2.2%

(1000 nm to 1630 nm)

±1.5%

(970 nm to 1630 nm)

±3.0%

(950 nm to 1630 nm)

±2.5%

(950 nm to 1630 nm)

Total uncertainty

2

±3.5% ±5 pW

(1000 nm to 1630 nm)

±2.8% ±5 pW

(970 nm to 1630 nm)

±5.0% ±500 pW

8

(950 nm to 1630 nm)

±4.5% ±500 pW

8

(950 to 1250 nm

max 23 dBm)

(1250 to 1630 nm

max 27 dBm)

Relative uncertainty:

7

- due to polarization

3

- spectral ripple

(due to interference)

4

≤ ±0.005 dB (typ. ±0.002 dB)

≤ ±0.005 dB (typ. < ±0.002 dB)

≤ ±0.005 dB (typ. ±0.002 dB)

≤ ±0.005 dB (typ. < ±0.002 dB)

Linearity (power):

5

- at 23 °C ±5 °C

- at operating temp. range

CW –70 dBm to +10 dBm, 1000 nm to 1630 nm

< ±0.02 dB

< ±0.05 dB

CW –50 dBm to +27 dBm, 950 nm to 1630 nm

< ±0.04 dB

8

< ±0.15 dB

8

Return loss

Typ. 60 dB

7

> 45 dB

> 47 dB

Noise (peak to peak)

6

< 5 pW

< 500 pW

Averaging time (minimal)

100 µs

100 µs

Analog output

Included

Included

Maximum safe input power

+16 dBm

+23.5 dBm (850 nm to 1650 nm)

+27.5 dBm (1250 nm to 1650 nm)

Dimensions (H x W x D)

57 mm x 66 mm x 156 mm

57 mm x 66 mm x 156 mm

Weight

0.5 kg

0.5 kg

Recommended recalibration period 2 years

2 years

Operating temperature

0 °C to 40 °C

0 °C to +35 °C

9

Humidity

Non-condensing

Non-condensing

Warm-up time

40 min

40 min

1. Reference Conditions:

• Power level 10 µW

(-20 dBm), continuous wave (CW)

• Parallel beam, 3 mm spot diameter on the

center of the detector

• Ambient temperature 23 °C ±5 °C
• On day of calibration (add ±0.3% for aging

over one year, add ±0.6% over two years)

• Spectral width of source ≤ 10 nm (FWHM)
• Wavelength setting at power sensor cor-

responding to source wavelength ±0.4 nm

2. Operating Conditions:

• Parallel beam, 3 mm spot diameter on the

center of the detector or connectorized fiber

with NA ≤ 0.2 (straight connector, options

C01/C02 also with angled connector ≤ 8°)

• For NA > 0.2 add 1%
• Averaging time 1 s

• Within one year after calibration,

add 0.3% for second year

• Zeroing required

3. All states of polarization at constant wave-

length (1550 nm ±30 nm) and constant

power, straight connector, T = 23 °C ±5 °C.

For angled connector (8°) add ±0.01 dB typ.
4. Conditions:

• Wavelength 1550 nm ±30 nm,

fixed state of polarization

• Temperature 23 °C ±5 °C
• Linewidth of source ≥ 100 MHz
• Angled connector 8°

5. Does not include noise; zeroing required
6. Averaging time 1 s, T = 23 °C ±5 °C,

ΔT ±1 °C, observation time 300 s.

Wavelength range 1200 nm to 1630 nm

7. Conditions:

• Wavelengths 1550 nm ±30 nm
• Standard single mode fiber, angled

connector min 8°

• With D-shape adapter 81001xx

return loss > 60 dB typical

8. For input power > +10 mW:

• Add typ. ±0.0016 dB/mW, or in case of

options C01/C02 for wavelength ≤ 1550 nm

add ±0.0006 dB/mW (guaranteed) using

adaptor Agilent 81000AF

• In case of decreasing power, allow

time for stabilization of the reading

(about 5 s per dB change)

• In case of decreasing power by more than

50 dB, allow recovery time of 3 minutes.

9. Max 30 °C above +20 dBm input power