Atec Agilent-81624B User Manual

Page 6

background image

6

Agilent 81623B

Agilent 81623B

Calibration option C85/C86

Agilent 81623B

Calibration option C01/C02

Sensor element

Ge, Ø5 mm

Wavelength range

750 nm to 1800 nm

Power range

–80 dBm to +10 dBm

Applicable fiber type

Open beam

Standard SM and MM max 100 µm core size, NA ≤ 0.3

Parallel beam max Ø4 mm

Uncertainty (accuracy) at

reference conditions

1

±2.2%

(1000 nm to 1650 nm)

±3.0% typ.

(800 nm to 1000 nm)

±2.2%

(1000 nm to 1650 nm)

±2.5%

(800 nm to 1000 nm)

±1.7%

(1000 nm to 1650 nm)

±3.0% typ.

(800 nm to 1000 nm)

Total uncertainty

2, 9

±3.5% ±100 pW

(1000 nm to 1650 nm)

±4.0% typ. ±250 pW

(800 nm to 1000 nm)

±3.5% ±100 pW

(1000 nm to 1650 nm)

±3.7% ±250 pW

(800 nm to 1000 nm)

±3.0% ±100 pW

(1000 nm to 1650 nm)

±4.0% typ. ±250 pW

(800 nm to 1000 nm)

Relative uncertainty:

- due to polarization

3

- spectral ripple (due to interference)

4

< ±0.01 dB

10

(typ. < ±0.005 dB)

< ±0.006 dB (typ. < ±0.003 dB)

Linearity (power):

5

at 23 °C ±5 °C

at operating temp. range

(CW –60 dBm to +10 dBm)

9

< ±0.025 dB

< ±0.05 dB

Return loss

7

> 50 dB, typ. > 55 dB

8

> 56 dB

Noise (peak to peak)

6

< 100 pW (1200 nm to 1630 nm)

< 400 pW (800 nm to 1200 nm)

Averaging time (minimal)

100 µs

Analog output

Included

Maximum safe input power

+16 dBm

Dimensions (H x W x D)

57 mm x 66 mm x 156 mm

Weight

0.5 kg

Recommended recalibration period

2 years

Operating temperature

0 °C to 40 °C

Humidity

Non-condensing

Warm-up time

40 min

1. Reference conditions:

• Power level 10 µW (-20 dBm),

continuous wave (CW)

• Parallel beam, 3 mm spot diameter on

the center of the detector

• Ambient temperature 23 °C ±5 °C
• On day of calibration (add ±0.3% for

aging over one year, add ±0.6% over

two years)

• Spectral width of source < 10 nm (FWHM)
• Wavelength setting at power sensor cor-

responding to source wavelength ±0.4 nm

2. Operating Conditions:

• Parallel beam, 3 mm spot diameter on the

center of the detector or connectorized

fiber with NA ≤ 0.2 (straight connector,

options C01/C02 also with angled

connector ≤ 8°)

• For NA > 0.2 add 1%
• Averaging time 1 s
• Within one year after calibration,

add 0.3% for second year

• Spectral width of source < 10 nm (FWHM)
• Wavelength setting at power sensor cor-

responding to source wavelength ±0.4 nm

3. All states of polarization at constant wave-

length (1550 nm ±30 nm) and constant

power, straight connector, T = 23 °C ±5 °C.

For angled connector (8°) add ±0.01 dB typ.
4. Conditions:

• Wavelength 1550 nm ±30 nm,

fixed state of polarization

• Temperature 23 °C ±5 °C
• Linewidth of source ≥ 100 MHz
• Angled connector 8°

5. Does not include noise; for wavelength

< 1000 nm applies for –50 dBm to +10 dBm
6. Averaging time 1 s, T = 23 °C ±5 °C,

ΔT ±1 °C, observation time 300 s
7. Conditions:

Wavelengths 1550 nm ±30 nm

Standard single mode fiber,

angled connector min 8°
8. With D-shape adapter 81001xx

return loss > 60 dB typical
9. For input power > 2 mW add ±0.004 dB/mW

(not for C01/C02); zeroing required
10. Specification valid for optical heads with

S/N starting with “DE413…”

(shipping began April 1, 2001)

Optical head specifications

All optical heads have to be operated with the single (Agilent 81618A) or dual (Agilent 81619A) Interface Modules