Agilent 81635A

Agilent 81634B

Sensor element

InGaAs (dual)

InGaAs

Wavelength range

800 nm to 1650 nm

800 nm to 1700 nm

Power range

–80 dBm to +10 dBm

–110 dBm to +10 dBm

Applicable fiber type

Standard SM and MM up to 62.5 μm core size,

NA ≤ 0.24

Standard SM and MM up to 100 μm core size,

NA ≤ 0.3

Uncertainty (accuracy) at

reference conditions

1

Typ. < ±3.5%

10

(800 nm to 1200 nm)

±3% (1200 nm to 1630 nm)

±2.5%

(1000 nm to 1630 nm)

Total uncertainty

2

Typ. ±5.5% ±200 pW

9, 11

(800 nm to 1200 nm)

±5% ±20 pW

8, 9

(1200 nm to 1630 nm)

±4.5% ±0.5 pW

(1000 nm to 1630 nm)

Relative uncertainty:

- due to polarization

3

- spectral ripple

(due to interference)

4

Typ. < ±0.015 dB

Typ. < ±0.015 dB

< ±0.005 dB

< ±0.005 dB

Linearity (power):

5

- at 23 °C ±5 °C

- at operating temp. range

CW –60 dBm to +10 dBm

Typ. < ±0.02 dB

9

(800 nm to 1200 nm)

< ±0.02 dB

9

(1200 nm to 1630 nm)

Typ. < ±0.06 dB

9

(800 nm to 1200 nm)

< ±0.06 dB

9

(1200 nm to 1630 nm)

CW –90 dBm to +10 dBm

< ±0.015 dB (1000 nm to 1630 nm)

< ±0.05 dB (1000 nm to 1630 nm)

Return loss

7

> 40 dB

> 55 dB

Noise (peak to peak)

6

Typ. < 200 pW (800 nm to 1200 nm)

< 20 pW (1200 nm to 1630 nm)

< 0.2 pW (1200 nm to 1630 nm)

Averaging time (minimal)

100 μs

100 μs

Analog output

None

Included

Maximum safe input power

+16 dBm

+16 dBm

Dimensions (H x W x D)

75 mm x 32 mm x 335 mm (2.8” x 1.3” x 13.2”)

75 mm x 32 mm x 335 mm (2.8”x 1.3”x 13.2”)

Weight

0.5 kg

0.5 kg

Recommended

recalibration period

2 years

2 years

Operating temperature

+10 °C to +40 °C

0 °C to +45 °C

Humidity

Non-condensing

Non-condensing

Warm-up time

20 min

20 min

1. Reference Conditions:

• Power level 10 μW (-20 dBm),

continuous wave (CW)

• Fiber 50 μm graded-index, NA = 0.2
• Ambient temperature 23 °C ±5 °C
• On day of calibration (add ±0.3%

for aging over one year, add ±0.6%

over two years)

• Spectral width of source < 10 nm (FWHM)
• Wavelength setting at power sensor must

correspond to source wavelength ±0.4 nm

2. Operating Conditions:

• Fiber ≤ 50 μm, NA ≤ 0.2
• 81634B only: connectors with 2.5 mm fer-

rule with flat face (fiber tip offset not more

than 0.3 mm from 2.5 mm cross-section)

with straight or angled polish

• 81635A only: for LC connector use 81000LI

adapter, not 81002LI; for MU connector

use 81000MI, not 81002MI

• Averaging time 1 s
• Only Agilent 81635A: For fiber 62.5 μm

graded-index (NA = 0.24): add ±2%

• Within one year after calibration,

add 0.3% for second year

• Operating temperature range as specified

humidity: none condensing

3. All states of polarization at constant wave-

length (1550 nm ±30 nm) and constant

power, straight connector, T = 23 °C ±5 °C.

For angled connector (8°) add ±0.01 dB typ.
4. Conditions:

• Wavelength 1550 nm ±30 nm, fixed

state of polarization, constant power

• Temperature 23 °C ±5 °C
• Linewidth of source 100 MHz
• Angled connector 8°

5. Does not include noise
6. Averaging time 1 s, T = 23 °C ±5 °C, ΔT ±1 °C,

observation time 300 s
7. Conditions:

• Wavelengths 1310 nm ±30 nm

and 1550 nm ±30 nm

• Standard single mode fiber,

angled connector min 8°

• T = 23 °C ±5 °C

8. For wavelengths > 1600 nm add ±0.06%/nm
9. For input power > 2 mW add ±0.02 dB
10. Add ±1% for wavelength 900 nm to 950 nm
11. Add ±3.5% for wavelength 900 nm to 950 nm