beautypg.com

3 conductivity system messages, Conductivity system messages – Xylem MULTI 3400 MULTI-PARAMETER METERS User Manual

Page 68

background image

What to do if...

Multi 3400i

68

Measured value too low

Measured value too

high

6.3

Conductivity system messages

Error message

OFL

OFL

OFL

OFL

Error message

E3

E3

E3

E3

Cause

Remedy

– Insufficient flow

– Provide flow to the probe

Cause

Remedy

– High amount of dissolved

substances

– Correct solubility function

using the salinity equivalent

– Air bubbles bump on the

membrane with high velocity

– Avoid direct flow to the

membrane

– The carbon dioxide pressure is

too high (> 1 bar)

– Measuring not possible

Cause

Remedy

The measured value lies outside
the measuring range

– Measuring cell not connected

– Connect measuring cell

– Cable broken

– Replace measuring cell

Cause

Remedy

– Measuring cell contaminated

– Clean cell and replace it if

necessary

– Unsuitable calibration

solution

– Check calibration solutions