beautypg.com

Xylem MULTI 3400 MULTI-PARAMETER METERS User Manual

Page 17

background image

Multi 3400i

Operation

17

Temperature measure-

ment in pH measure-

ments

You can perform pH measurements with or without a temperature sen-
sor as well as with the temperature sensor of an oxygen sensor or a
conductivity measuring cell. The measuring instrument recognizes
which sensors are connected and switches automatically to the correct
mode for the temperature measurement.

The following cases are distinguishable.

* If you do not wish that, you can:

– either disconnect the 2nd sensor and use the manual temperature

input or

– use an electrode with a temperature sensor.

If a temperature sensor is connected, it is indicated on the display by
TP.

Note
When calibrating without a temperature sensor (no TP display indicator
displayed), enter the current temperature of the respective buffer solu-
tion manually using the <

> <> keys while keeping the

TER> key depressed.

Temperature sensor

Display

Mode

pH

Cond or
Oxi

yes

-

TP

Automatic with
pH temperature sensor

yes

yes

TP

-

-

Manual

-

yes

TP
flashes

The temperature value of the
second probe (Cond or Oxi) in
the same sample is taken
over for the pH measurement*