6 salt content correction, Salt content correction, Section 4.3.6 s – Xylem MULTI 3400 MULTI-PARAMETER METERS User Manual
Page 33: Content, Correction

Multi 3400i
Operation
33
4.3.6
Salt content correction
A salt content correction is required in the oxygen concentration mea-
surement of samples with a salt content of more than 1 g/l. To do this,
you have to enter the salinity equivalent of the test sample (range 0.0 -
70.0) and to switch on the salinity correction.
Note
With the Multi 3400i handheld meter, you can measure the salinity.
How to proceed is described in section 4.4.3 M
Entering the salt con-
tent
Note
How to switch on the salt content correction is described on page 27.
1
Determine the salinity of the test sample (any method, see also
section 4.4.3 M
2