beautypg.com

5 d. o. calibration, D. o. calibration – Xylem MULTI 3400 MULTI-PARAMETER METERS User Manual

Page 30

background image

Operation

Multi 3400i

30

4.3.5

D. O. calibration

Why calibrate?

D. O. probes age. This changes the slope of the D. O. probe. Calibra-
tion determines the current slope of the probe and stores this value in
the instrument.

When to calibrate?

z After connecting another D. O. probe
z When the probe symbol flashes (after the calibration interval has ex-

pired).

Calibration procedure

The calibration is performed in water vapor-saturated air. Use the
OxiCal

®

-SL air calibration vessel (accessories) for the calibration.

AutoRead

The calibration procedure automatically activates the AutoRead func-
tion. The AR display indicator flashes. The calibration process is fin-
ished when AR stops flashing.

Displaying the calibra-

tion data

You can view the data of the last calibration on the display. The pro-
ceeding is described on page 48.

Printing the

calibration protocol

The calibration record contains the calibration data of the current cali-
bration. You can transmit the calibration protocol to a printer via the se-
rial interface (see page 51).

Note
You can automatically print a calibration record after the calibration. To
do so, connect a printer to the interface according to section 4.7.2 be-
fore calibrating. After a valid calibration, the record is printed.

Sample printout:

CALIBRATION PROTOCOL
02.03.01 14:19
Device No.: 12345678
CALIBRATION 02
Cal time: 02.03.01 / 14:19
Cal interval: 14d
OxiCal Tauto AR
Relative Slope: 0,88
Probe: +++